The tibial pin guide (in addition to the product packaging
label) will have patient specific identifiers: Patient Name,
Lot No., Size and Patient Anatomy (R/L). Verify the
accuracy of these identifiers prior to opening the sterile
package (Figure 1).
Note:
The size information was selected pre-
operatively based on the Patient Proposal. Final
implant sizing may change due to intra-operative
assessment of implant fit and/or joint gap balance.
Prior to use, insert the TRUMATCH Solutions Drill Guides
(Part Number (P/N) 2004-20-925) into the two anterior
openings of the plastic tibial pin guide by twisting in a
clockwise direction until tightened (Figure 2).
Note:
The TRUMATCH Solutions Drill Guides (P/N
2004-20-925) are reusable after sterilization. A
minimum of four (4) drill guides should be on hand
for a case. They are shipped separately from the
TRUMATCH Solutions Pin Guides.
For optimal handling and placement stability of the tibial
pin guide, first insert the HP Extra Medullary Tibial Uprod
(P/N 9505-01-228) into the anterior holes of the tibial pin
guide. Then slide the Rod Extension (P/N 2004-20-923)
over the distal end of the uprod. This will lengthen it to
reach the patient’s ankle. Grasp the guide using the
medial and lateral finger pads (Figure 3A). Do not grasp
the uprod or the area on which the metal drill guides are
located. (Figure 3B).
Figure 2
Figure 1
Figure 3A
Figure 3B
PROXIMAL TIBIAL RESECTION
4 DePuy Synthes Joint Reconstruction TRUMATCH Personalized Solutions Pin Guides Surgical Technique