The Arise LAL is divided into three distinct 
anatomical zones: head, seat, and foot. Each zone 
is adjusted to ensure optimal pressure relief and 
provide a comfortable sleeping surface. 

Maximum pressure relief is achieved through 

cell and allowing controlled amounts of air to 
escape, thus equalizing the pressure between 
the patient and the therapy cells. This distributes 
the patient’s weight evenly over a wide surface 
area resulting in average pressure readings below 
capillary closure levels.

Patients are at risk for skin maceration if excess 
moisture is permitted to accumulate beneath 
the patient. This may be due to perspiration, 
incontinence or wound drainage. 

On the Arise LAL, moisture is controlled via the 

coated nylon therapy pad. The moisture vapor 
permeable fabric of the therapy pad allows a 

pad and wicks away excess moisture.

Shearing occurs when the skin is stationary 
in relation to the support surface, while the 
underlying tissues and vessels are stretched 
and damaged. When a patient’s skin rubs 
against another surface, the result is friction. 
The top surface of the Arise LAL therapy pad is 
constructed from a very smooth nylon fabric with 
low friction and shear properties to protect the 
patient’s skin from these damaging forces.

The Arise LAL

®

, provided by Joerns Healthcare, is a 

unique therapy system that provides pressure relief 
by combining low air loss with pulsation. Low air loss 
therapy has been demonstrated to reduce the risk of 
pressure ulcers as well as being a valuable aid in the 
treatment of pressure ulcers. 

Warning:

 The risk of entrapment can develop when 

equipment is placed on bed frames that leave gaps 
of even a few inches between the mattress and the 
headboard, footboard, and bed or side rails. The 
equipment is NOT to be used when such gaps are 
present. See “Important Precautions” section of   
this manual.

The Arise LAL mattress replacement systems and 
overlays are suitable for both the treatment of 
existing pressure ulcers Stage I through Stage IV 
as well as those who have been assessed at risk 
from the complications of immobility. The Arise LAL 
is quiet, comfortable and simple enough for single 

allow for easy adjustment of patient comfort. The 
Arise LAL offers special features to increase peace 
of mind such as a power interruption alarm and a 
patient position sensor which optimizes support for 
seated patients (Fowler Boost).

Additionally, low friction and shear materials, 
together with average interface pressures below 
capillary closure levels (32mm of Mercury),  
means that the Arise LAL meets the comfort and  
clinical requirements.

We have ensured that the Arise LAL addresses the 
four key areas in the treatment of compromised skin: 
pressure relief, moisture control, and reduction in 
both friction and shearing forces.

Electronic Copy: Reference Only

Notice: Document considered current at time of printing. Verify revision before use. 

Содержание Arise LAL 1000

Страница 1: ...To avoid injury read user s manual before using Electronic Copy Reference Only Notice Document considered current at time of printing Verify revision before use...

Страница 2: ...properly within the protrude over the side or between the bed rails when the mattress is being used Danger Explosion Hazard Do not use in the in the presence of smoking materials or open support combu...

Страница 3: ...uidelines intended to reduce the risk of entrapment including dimensional limits for critical gaps and spaces between bed system components and clinical guidance for assessment and implementation of b...

Страница 4: ...Power Failure CairRails Optional Accessories Troubleshooting 13 Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures 13 14 Recommended Linen Changing the Therapy Pad Patient Pos...

Страница 5: ...ssure relief by combining low air loss with pulsation Low air loss therapy has been demonstrated to reduce the risk of pressure ulcers as well as being a valuable aid in the treatment of pressure ulce...

Страница 6: ...for these patients should be considered on an individual basis and discussed with the attending physician The Arise LAL is comprised of two components Therapy control unit Therapy mattress system per...

Страница 7: ...ed low pressure heel section provides lower interface pressures Durable base tub is constructed from 100 heavy weight 1680 denier nylon with a 1 5 oz urethane coating and incorporates bed attachment l...

Страница 8: ...poor installation of a ground wire that causes the loss of proper ground at the outlet Inadequate grounding at electrical outlets can occur even if there is a ground wire Wires can become loose over t...

Страница 9: ...eat section securely to the sleep deck of the bed frame If the therapy pad is not already on the overlay attach it securely using the zippers Hang the control unit on the foot of the bed facing away f...

Страница 10: ...setting Refer to Comfort Adjust Setup section Figure 2 on the lower right of the control CairRails by moving the control knob to the up If the patient needs the extended width functionality of the be...

Страница 11: ...d Pulsate Static The unit starts in the Static mode which is the standard low air loss therapy Pulsate Pulsate will slightly decrease the pressure in all cushions every 30 seconds then return to the p...

Страница 12: ...nd visual alarm Power Interruption Brown out In the event of a brief power interruption up to twenty 20 minutes the Arise LAL control unit will automatically restart and reset to the previous Comfort...

Страница 13: ...ly Special linens are not necessary for the Arise LAL While there is no need for a bottom sheet the therapy pad should be covering the therapy cells at all times The patient should never be lying dire...

Страница 14: ...should always be in the lowest practical position when the patient is unattended Make sure areas under and around the frame are clear of objects persons and parts of body before adjusting height Warni...

Страница 15: ...y pad can be laundered in a washer and dryer with warm water no more than 120 Fahrenheit A non bleach detergent should be used sparingly Wipe dry or allow to air dry Do not use any steam cleaning devi...

Страница 16: ...LAL 1000 Mattress 48 122 cm W 80 203 cm H 8 5 22 cm D Arise LAL 1000EX Mattress 36 42 48 91 107 122 cm W 80 88 203 224 cm L 8 5 22cm D USA 120V AC 60 Hz 4A less than 1A running Operating Conditions A...

Страница 17: ...Electronic Copy Reference Only Notice Document considered current at time of printing Verify revision before use...

Страница 18: ...e instructions is not covered Joerns Arise LAL contains various parts that wear from normal use Joerns Healthcare s obligation under it s warranty is limited to supplying replacement parts servicing o...

Отзывы: