DIRECTIONS FOR USE
EN-10
NORMAL MEASUREMENT
1.
Connect each accessory to the device
1) Remove the connecting tube port cap from the device.
2) Connect one side of the connecting tube connector securely to the
connecting tube port and the other side of the connecting tube
connector securely to the Tongue Pressure Probe.
2. Turn on
Press the power button on the device to turn it on.
The LCD screen lights with a buzzer sound and after all indications are displayed, the unit
automatically enters the standby mode.
<Power OFF> <FULL Display> <Standby mode>
3. Adjust the internal pressure automatically
Make sure the device is in standby mode and press the Prep/Reset button.
The internal pressure is automatically adjusted and zero-point correction is performed.
<Operation of the device>
1.
When the Prep/Reset button is pressed, the pressurization pump operates with a beeping sound, and the internal pressure
adjustment starts. At this time, the
“Wait” icon on the LCD screen flashes and the “PTP” value increases. The Level sign indication will
also increase to the right. When the total number of the Level sign indication reaches 10, it indicates that the set value of internal
pressure has reached 19.6 kPa.
2.
When the inside of the connected the device and accessories is adjusted to 19.6 kPa,
“MTP”, “PTP” and Level sign indication are
automatically reset and
“Measure” icon lights. In this status, it is ready for the measurement.
<Standby mode>
<Adjusting internal pressure> <Ready for the measurement>
Measurement preparation
Hold only the grips of the Tongue Pressure Probe while adjusting the internal pressure. Touching the
balloon or tube at this time may cause incorrect internal pressure adjustment resulting in
measurement errors or impossibility.
CAUTIONS
Screw without gap