5. 

Place the wedge on the inner probe holder arm’s button and clamp in place with the outer 

probe holder arm 

(Fig. 4)

. Tighten the probe holder arm adjustment knob.

4.2.  Encoder Setup 

1. 

Loosen the encoder’s thumb screw and lower the encoder wheel’s pivot joint towards the 

scan surface 

(Fig. 5)

.

2. 

Tighten the encoder thumb screw and ensure adequate spring tension of the encoder 

wheel to the scan surface 

(Fig. 6)

.

TROUBLESHOOTING

Encoder not 
incrementing

Encoder wheel not in contact with 
scan surface

Rotate pivot joint until wheel is in 
contact with scanning surface and 
spring joint is slightly depressed. 

(see 4.2. Encoder Setup)

Encoder connector not properly 
connected to scanning device

Check scanning devices instructions 
to properly connect an encoder.

SPARE PARTS

6

5

8

9

10

7

4

1

2

3

Part #

Description

1

CKA010

ODI-II Case

2

3

4

5

6

7

8

9

10

BOM ID

CKS015

Encoder Clamping Mount

CKS011-X-05 Encoder Assembly 

(Various encoder connector types)

PH0082

Probe Holder Arm Adjustment Knob

CK0018

Clamp Arm

 (50 mm, Various sizes available)

CKS014

Yoke Clamping Mount

PH0011-X

Pivot Button

 (Various available)

PH00XX

MD050-010

Arm Clamp Screw, SHCS, M4x0.7 X 8 mm, SST

CK0047-X

Dovetail Slide Frame Bar

 (30 mm, Various sizes available)

Arm Style 

 (Various available)

Visit jireh.com for

information

Fig. 4

 - Mount wedge

Fig. 5

 - Lower encoder wheel

Fig. 6

 - Ensure spring tension

Отзывы: