25/25
6.2 Accuracy
The following indicators test conditions:
Temperature: 20
±
3
℃
Humidity: <80% RH
Warm-up time of more than 15 minutes
Calibration time within 1 year
Voltage output:
Voltage output range
25V~1000V DC
Voltage output accuracy
1%
±
2V
Voltage resolution
1V
Voltage readback accuracy
2%
±
1V
Maximum charging current
1.8mA
Short-circuit test current
2mA
Range display range:
Test voltage
Resistance range
Display range
(Ω)
Resolutio
(Ω )
25V
≤
V
<
100V
2M
Ω
0.000~4.000M
0.001M
20M
Ω
1.90M~40.00M
0.01M
200M
Ω
19.0M~400.0M
0.1M
100V
≤
V
<
500V 2M
Ω
0.000~4.000M
0.001M
20M
Ω
1.90M~40.00M
0.01M
200M
Ω
19.0M~400.0M
0.1M
2000M
Ω
190M~4000M
1M
500V
≤
V
<
1000V
2M
Ω
0.000~4.000M
0.001M
20M
Ω
1.90M~40.00M
0.01M
200M
Ω
19.0M~400.0M
0.1M
4000M
Ω
190M~9990M
1M
Resistance measurement accuracy:
Test voltage
Range
Basic accuracy
25V
≤
V
<
100V
0.000 M
Ω
to 2.000 M
Ω
±
2% rdg.
±
5 dgt.
1.90 M
Ω
to 20.00 M
Ω
±
2% rdg.
±
5 dgt.
19.0 M
Ω
to 200.0 M
Ω
±
5% rdg.
±
5 dgt.
100V
≤
V
<
500V
0.000 M
Ω
to 2.000 M
Ω
±
2% rdg.
±
5 dgt.
1.90 M
Ω
to 20.00 M
Ω
±
2% rdg.
±
5 dgt.
19.0 M
Ω
to 200.0 M
Ω
±
5% rdg.
±
5 dgt.
500V
≤
V
≤
1000V
0.000 M
Ω
to 2.000 M
Ω
±
2% rdg.
±
5 dgt.
1.90 M
Ω
to 20.00 M
Ω
±
2% rdg.
±
5 dgt.
19.0 M
Ω
to 200.0 M
Ω
±
2% rdg.
±
5 dgt.
190 M
Ω
to 4000 M
Ω
±
5% rdg.
±
5 dgt.
4010 M
Ω
to 9990 M
Ω
±
25% rdg.