BR1 User Manual
29
M-BR1-001-01
10
SPECIFICATIONS
Table 7: BR1 optical and electrical specifications sheet
Parameter
Specification
Single-mode
Multimode
Fiber Type (μm)
9/125
50/125 or 62.5/125
Encircled Flux Standard
N/A
IEC 61280-4-1
Operating Wavelengths (nm)
1310 / 1490 / 1550 / 1625 /
1650
850 / 1300
Backreflection Range (dB)
0 to -85
0 to -60
Return Loss Accuracy (dB)
1,2
± 0.4
Detector Type
2mm InGaAs / 5mm Ge / Integrating Sphere
Powe Range (dBm)
0 to -80 / 0 to -60 / o to -40
Insertion Loss Accuracy (dB)
± 0.03 (< 5 dB loss)
±0.15 (> 5 dB loss)
Absolute Power Accuracy (dB)
3
± 0.25
Remote Interface
USB/Ethernet
Display
5” touchscreen
Input Voltage
100 to 240 V AC, 50 to 60 Hz
Power Consumption (VA)
60 maximum
Notes:
1
add ± 0.1 dB for every 1 dB below -60 dB (single-mode)
2
add ± 0.1 dB for every 1 dB below -45 dB (multimode)
3
Measured at -10 dBm