BR1 User Manual
25
M-BR1-001-01
8
MAINTENANCE
Warning
•
Devices with malfunctioning lasers must be returned to the manufacturer for
repair.
Cleaning the Unit
1.
Unplug the unit from the line power.
2.
Clean the enclosure with a damp cloth.
3.
Do not plug the unit back in until it is completely dry.
Cleaning the Output
Warning
•
Connecting contaminated or damaged connectors to the BR1 can damage the
unit and affect its performance.
•
Damaging the output fiber during maintenance can affect the performance of
the unit.
1.
Inspect all connectors before each mating and if needed, clean with a lint-free wipe and/or IPA.
Figure 17 shows a dirty connector requiring cleaning. Figure 18 shows a clean connector ready to
be mated.
2.
Loosen the front panel thumbscrews.
3.
Gently remove the output panel. Ensure a clear line of sight to the fiber to prevent any stress on
the output fiber.
4.
Remove the connectors from the mating sleeves. For multiple output
BR1’s
, a marking is visible
on the fiber to distinguish which fiber is used for which output.
5.
Clean the connectors and mating sleeves in accordance with the section Cleaning Jumper
Connectors on page 27.
6.
Reinstall the connectors into the mating sleeves.
7.
Reinstall the output panel with the thumbscrews. To avoid damaging the fibers, keep a clear line
of sight to the fiber as it spools back inside the unit. It should lay flat and spool back in without