Measurement and instrumentation
Multimeter model UT60A
Réf :
291 033
ENGLISH
31
To test the fuse:
1. Set the rotary switch to
Ω
and select
by pressing
BLUE
button.
2. Plug a test lead into the terminal HzV
Ω
and touch the probe tip to the
10A or µA mA terminal.
3. If the Multimeter beeps, the fuse is good.
4. If the dis play shows
OL,
replace the fuse and test again.
5. If the display shows any other value, have the multimeter serviced and
contact your dealer immediately.
If the Multimeter does not work while the fuse is all right, send it to your dealer
for repair.
7.3 Replacing the Fuses
Warning
To avoid electrical shock or arc blast, or personal injury or damage to
the multimeter, use specified fuses ONLY in accordance with the
following procedure.
To replace the multimeter's fuse:
1. Press the POWER to turn the Multimeter off and remove all
connections from the terminals.
2. Remove the screw from the battery compartment, and separate the
battery compartment from the case bottom.
3. Remove the 2 rubber feet and 2 screws tram the case bottom, and
separate the case top from the case bottom.
4. Remove the fuse by gently prying one end loose, and then take out
the fuse from its bracket.
5. Install ONLY replacement fuses with the identical type and
specification as follows and make sure the fuse is fixed firmly in the
bracket.
Fuse 1: 0.5A, 250V, fast type Glass fuse, Ø5x20mm.
Fuse 2: 10A, 250V, fast type Glass fuse, Ø5x20 mm.
6. Rejoin the battery compartment and the case top, and reinstall the
screw.
7. Rejoin the case bottom and case top, and reinstall the 2 screws and 2
rubber feet.
Replacement of the fuses is seldom required. Burning of a fuse always results
from improper operation.
7.4 Replacing the Battery
Warning
To avoid false readings, which could lead to possible electric shock or
personal injury, replace the battery as soon as the battery indicator
appears.
To replace the battery:
1. Press the
POWER
to turn the Multimeter off and remove all
connections from the terminals.
2. Remove the screw from the battery compartment, and separate the
battery compartment from the case bottom.
3. Remove the battery from the battery compartment.