Measurement and instrumentation
Multimeter model UT70A
Réf:
291 034
ENGLISH
14
Remarks:
Frequency response 40 – 400 Hz showing sine wave effective
value (mean value) at “10A Range” test no more than 10sec. at
interval of 15min.
Resistance
Range
Resolution
Accuracy
200
Ω
0.1
Ω
± (0.8 % + 3)
2 k
Ω
1
Ω
20 k
Ω
10
Ω
200 k
Ω
100
Ω
2 M
Ω
1
k
Ω
± (0.8 % + 1)
20 M
Ω
10
k
Ω
± (1 % + 5)
2000 M
Ω
1
M
Ω
± [5 % (reading – 10) +10]
Input protection: 500 Vrms.
Remarks:
it is normal as test response slow at 2000M
Ω
.
Capacitance
Range
Resolution
Accuracy
Input protection
20 nF
10 pF
200 nF
100 pF
2 µF
1 nf
± (2.5 % + 5)
100 µF
100 nF
± (5 % + 4)
250 Vrms
Inductance
Range
Resolution
Accuracy
Input protection
2 mH
1 µH
20 mH
10 µH
200 mH
100 µH
± (2 % + 10)
20 H
10 mH
± (3 % + 10)
250 Vrms
Tested inductance: Q>10, internal impedance<1.3 k.
Auto range frequency test
Range
Resolution
Accuracy
Input protection
2 kHz – 10MHz
1 Hz (MIN)
± (0.1 % + 3)
500 Vrms
Remarks:
Input sensitivity<0.8 Vrms.
Transistor hFE measurement
Range
Resolution
Accuracy
Input protection
hFE
1ß
hFE
approximation
(0-1000
β
) is
shown for the
tested transistor
(NPN, PNP)
1 bo
≈
10 A
Vce
≈
2.8 V