Measurement
Air/Water oxymeter Initio
Ref :
211022
ENGLISH
8
4 Adjustment
The first regulation of the oxymetre is only made on the range of
measure 0-25 %, it is valid for the measure in the air and in the water.
4.1
Period of stoking
Wait until the 5 minutes
heating
period
is over (the DEL of state of stoking
does not flash any more).
4.2 Calibration
The probe in the air, press the button of calibration.
CAL is displayed on screen for a few seconds and the red LED flashes.
.
Wait for some moments. In the stop of the blinking, your oxymètre is ready to
work, you are 20.9 %.
The oxymètre is so reglé to work in the air(sight) and in the water.
Should you want to use the oximeter in water immediately after the initial
adjustment, allow 5 to 8 minutes for the probe to stabilize in the liquid
medium.
5 Measurements
The calibration is stored for several days. Hence it is pointless to repeat this
operation each time the oximeter is used.