NOTE: Be sure you have the transmitter loop #1 connected to a load of 250 to 500 Ohms. A unit installed in a control loop 
is a good example of this loop load. You may also use a load resistor in the range of the above value. 

5. Press the black and white "I/O" button on the HART terminal. The terminal will go into self-test, then into the main 
screen. If not connected properly, you will get a "No device found" message. 

6. From the main screen, press keypad key #1, "Device Setup". 

7. From the "Device Setup” screen, press key #3, "Basic Setup". 

8.  Press key #3.  You are now in "Range Values" screen. 

To set low value 

To set the low value (4 mA), select key #1, PV LRV (Process Variable, Lower Range Value). You are now in the PV LRV 
screen. The current low value is displayed. Below this value is a highlighted value. Key in the desired low value (example 
3.00 in. is shown; if 4 inches is desired, key in 4.) When the new desired low value is keyed in, press the "enter" (F4) 
button located below the LCD display, right. To write the changed lower value to memory, press the "SEND" key now. 
Next you will see two "WARNING" screens that ask if you are sure. If your new low values are correct, press "OK" for both 
messages.  This action resets the Lower Range Value, or 4 mA position into the transmitter's memory.  Go back to the 
"Range Values" screen to verify that the new parameters have been accepted into the transmitter memory. 

You may now exit program mode or continue on to reset the upper value. If you choose to exit the program mode, replace 
the calibration jumper to the "ON” position, and return your controllers to automatic. 

To set high value 

You should now be in the "Range Values" screen. To set the 20 mA (Upper Range), press key #2. You are now in the "PV 
URV" (Process Variable, Upper Range Value) screen. As in the lower value screen, the current value is displayed with a 
highlighted number below it. To change the upper value, key in the desired value. You may use whole numbers or whole 
numbers and decimal numbers (40 = 40 inches, or 40.5 = 40.50 inches). Whole numbers will be entered as their decimal 
equivalents by HART automatically. Key in the desired upper range value desired. Press the "Enter" (F4) button. 

You are back in the "Range Values" screen. If the numbers for lower and upper are correct, press the "Send" key. You will 
get a "WARNING!" Press the "OK" button. You will again get "WARNING!" Press ”OK" again. 

Startup is now complete. 

CAUTION!  

Do not enter a high value that exceeds the active length of the sensor! 

7.2 Calibration 

The MTII4200 has three buttons for calibration: 

   Enter 

 

Pressing any button will cause the unit to enter calibration. 

   Down Arrow 

Pressing all three buttons at any one time will exit the calibration 

   Up Arrow  

mode. 

 

1.)  Turn the power on to the unit. 
2.)  Wait for the self-diagnostics to complete. 
3.)  Press any button to start calibration mode. 
4.)  The screen will display “Cal Lev 1?”. 
5.)  The far right button is the “Enter” key. 
6.)  Press the Enter key. 
7.)  The screen will display “Set Zero?”. 
8.)  Move the float to the zero, (4mA), location. 
9.)  Press the Enter key. 

10.)  The screen will display “Zero (corresponding number)”. 
11.)  Press Enter to reset the (corresponding number) to zero and hold for approximately 1-2 seconds. 
12.)  The screen will display “Accept?”. 
13.)  Press Enter (You have now set the zero point). 
14.)  The screen will display “Cal Lev 1?”. 
15.)  Press Enter. 
16.)  The screen will display “Set Zero?”. 
 

Содержание MTII4200 Series

Страница 1: ...D B TOP RIGHT HAND C BOTTOM RIGHT HAND D BOTTOM LEFT HAND ALARM SETTING L 3 85 mA DEFAULT H 22 5 mA MAX SENSING LENGTH IN INCHES ON 6 INCH INCREMENTS LCD DISPLAY M mA OUTPUT DEFAULT I INCHES P PERCENT...

Страница 2: ...II4200 transmitters are available in a single cavity explosion proof enclosure 2 1 MTII4200 Transmitter Specifications PARAMETER SPECIFICATIONS LEVEL OUTPUT Measured Variable Liquid level interface le...

Страница 3: ...f Operation The MTII4200 transmitters precisely sense the position of an external float by applying an interrogation pulse to a waveguide medium This current pulse causes a magnetic field to instantly...

Страница 4: ...Transmitter Always follow applicable local and national electrical codes and observe polarity when making electrical connections Never make electrical connections to the MTII4200 transmitter with powe...

Страница 5: ...how the HART protocol is applied to the MTII4200 level transmitter only Using the HART interface allows for calibration without having to remove the transmitter from the process and position the float...

Страница 6: ...to exit the program mode replace the calibration jumper to the ON position and return your controllers to automatic To set high value You should now be in the Range Values screen To set the 20 mA Upp...

Страница 7: ...n until it is Press the enter button to set the low level of the range At this time position the float magnet at the desired low level point and keep it there while pressing the Enter button and holdi...

Отзывы: