18
351 050/REV B/09-06
12
3.6.4 Temperature units
The display will show:
The secondary display will show UNITS and will then scroll TEMPERATURE UNITS after 10
seconds.
Select °C or °F using the Up/Down arrows.
The Left/Right arrow keys and the CAL/CLR key have no function during this set up.
Symbol will flash while adjustment is being made.
To accept °C or °F, press the STO key. The symbol will disappear and the secondary display will
momentarily show SAVED and the next set up menu option will be displayed.
To exit the set up menus press the ESC key at any time. This will return the instrument to the
MODE menu.
3.6.5 Manual temperature
If the ATC probe is not connected to the instrument the display will show:
The secondary display will show MAN °C (or °F) and will then scroll MANUAL TEMPERATURE °C
(or °F).
Measure the temperature of the buffer solutions(s) to be used and adjust the displayed reading to
these values. Adjustment of the flashing digit can be made using the Up/Down arrow keys. The
Left/Right arrow keys are used to select the next adjusting digit which will flash when selected.
Once the reading has been adjusted to the correct value, press the STO key. The instrument
display will update and show the Baud Rate set up menu.
Any changes to manual temperature settings reflect on 3.6.4.
To exit the set up menus press the ESC key at any time. This will return the instrument to the
MODE menu.