3.1 Theory of pH Measurement

pH is a unit of measurement which defines the degree of acidity or alkalinity of a solution. It is
usually measured on a scale of 0 to 14. The pH value quantifies the degree of hydrogen ion activity
of an acid or a base.

The internationally accepted symbol, pH, is derived from “p”, the mathematical symbol of the
negative logarithm and “H”, the chemical symbol for Hydrogen. The pH value is the negative
logarithm of Hydrogen ion activity as shown in the mathematical relationship pH=-log[H

+

].

The pH value of a solution is directly related to the ratio of the hydrogen ion [H

+

] and the hydroxyl ion

[OH

-

 ] concentrations. If the concentration of H

+

 is greater than OH

-

, the material is acidic and has a

pH value of less than 7. Conversely, if the concentration of OH

-

 is greater than H

 the material is

basic, with a pH value greater than 7. If the concentrations of H

+

 and OH

-

 are equal the material is

neutral with a pH value of 7.

It can, therefore, be seen that pH is a measurement of both acidity and alkalinity, even though by
definition it is a selective measurement of hydrogen ion activity. The logarithmic relationship
between hydrogen ion concentration and the pH unit means that a change of one pH unit
represents a ten-fold change in hydrogen ion concentration.

3.2 pH Measurement

pH can be measured by using either pH papers/indicators or a pH meter, dependent on the level of
accuracy required. pH papers or indicators change colour as the pH level varies. These can be
used as a guide to the pH colour level, but can be limited in accuracy and difficult to interpret
correctly in murky or coloured samples.
For greater accuracy the use of a high impedance pH meter is recommended, together with a pH
measuring electrode and reference electrode.

Each component part of the measurement system can be described as follows:
a)

the pH meter - is a high impedance amplifier used to accurately measure the minute

electrode voltages produced. The pH meter will display the results directly in pH units on either an
analogue or digital display. Voltages can also be read for special applications, ORP (Oxidation-
Reduction Potential) measurements or with Ion Selective Electrodes.

b)

the pH electrode - is a hydrogen ion sensitive glass bulb, with a millivolt output that varies with

the changes in the relative hydrogen ion concentration inside and outside of the bulb. The pH
electrode has very high internal resistance, making the voltage change with pH difficult to measure.
The input impedance of the pH meter and leakage resistances are therefore important factors.

c)

the reference electrode - these cells consist of an internal element, usually a silver/silver

chloride wire, electrolyte (KCl) and a liquid junction. The liquid junction provides a leak path for the
internal electrolyte to “weep” into the sample chamber and provide an electrical contact with the
liquid to be measured. If the liquid junction is inefficient then measurement will be inaccurate. It is
common for the reference electrode to be incorporated into the pH electrode. It is then called a
combination electrode. The Model 3505 is supplied with a combination electrode.

350 550/REV A/03-03

5

Содержание 3505

Страница 1: ...350 550 REV A 03 03 Model 3505 pH mV Temperature Meter Operating Manual...

Страница 2: ...and service personnel employ a safe system of work in addition to the detailed instructions specified in this manual 3 References should always be made to the Health Safety data supplied with any chem...

Страница 3: ...rement 3 1 pH Measurement 3 2 Preparation of Buffer Solution 3 3 Solution Temperature Values 3 4 Good Practice Guidelines 3 5 pH Calibration Using Buffers 3 6 Performing Measurements 3 7 Section 4 Mai...

Страница 4: ...3505 has a liquid crystal display and is battery operated The use of an optional power supply permits the unit to be operated from the a c mains supply 1 2 Instrument Specifications Range Resolution 2...

Страница 5: ...ing plug 009 146 7 Battery PP3 8 Condensed operating instructions 350 551 9 Instruction manual 350 550 Any shortages or damage should be reported immediately to the Manufacturer or your local Distribu...

Страница 6: ...lance out the electrode offset between the pH measurement cell and reference This should be set with the electrode immersed in pH7 buffer This control adjusts the electrode effect between 60mV to 60mV...

Страница 7: ...ch allows combination pH or redox electrodes to be used 3 Temp Socket 8 pin mini DIN socket This allows the Automatic Temperature Compensation ATC probe 027 500 to be connected 4 Analog Out 2 x 4mm so...

Страница 8: ...ies These can be used as a guide to the pH colour level but can be limited in accuracy and difficult to interpret correctly in murky or coloured samples For greater accuracy the use of a high impedanc...

Страница 9: ...te and repeatable results it is essential to follow the manufacturers instructions carefully 3 4 Solution Temperature Values The value of all buffer solutions varies with solution temperature For accu...

Страница 10: ...gel filled electrodes are not refillable 5 DO NOT touch the sensitive glass pH membrane or reference junction during use Excess droplets of solution may be removed by gently blotting with filter paper...

Страница 11: ...he controls are interdependent 5 Rinse the electrode s in deionised water 6 Select pH on the Function switch When this position is selected the calibration settings are locked no further adjustment ca...

Страница 12: ...socket on the rear panel b Select C on the Function switch The display will show ATC probe temperature directly in C 3 Temperature Measurement Manual a If Manual temperature compensation is being use...

Страница 13: ...nd in the solution until clean Rinse and repeat as necessary Electrodes which have been allowed to dry out should be soaked overnight in warm distilled water Table of Cleaning Agents for Glass Electro...

Страница 14: ...died combination 12mm diameter For liquids 924 005 General purpose glass bodied combination 12mm diameter For liquids For a complete listing of all available electrodes please contact your local distr...

Страница 15: ...ted thermometer Slope Buffer controls 3505 incorrectly set Set Dial to pH Cal not working Battery symbol is displayed Battery running out Less than 8hrs lifetime left replace battery If the above does...

Страница 16: ...ture input check Remove the temperature probe and apply a 10Kohm resistor across the pins of the temp input as described in fig 7 2 1 Fig 6 2 1 Temperature input with connection detail 350 550 REV A 0...

Страница 17: ...dard EN 50082 1 1992 Electromagnetic compatibility Generic immunity standard Performance criterion B EN 61010 1 1993 Safety requirements for electrical equipment for measurement control and laboratory...

Страница 18: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: