background image

CAL

innoCon 6800 Monitor Manual

Page 15

Saturation Calibration

Concentration Calibration

Please refer to "

Saturation Calibration

Current :

Current nA value of the electrode

Current nA value of the electrode

Current nA value of the electrode
Current saturated air temperature

Input Temp.:
Input Sat.:

Input Stand Saturation value

and Press ENTER

+80.0nA

25.0℃

100%

1

1

Wash the electrode dry and put it into saturated air, input the current saturated air temperature 

the next interface.

    

Saturation Calibration

Current :

Input Temp.:
Input Sat.:

Waiting for reading stable

and then press ENTER key

to exit

+80.0nA

25.0℃

100%

1

Store calibration value

1

CAL

→ YES

→ NO

Concentration Calibration

1

Restore

Restore

1

→ YES

→ NO

www.jensprima.com

Current :

Input Temp.:
Input Conc.:

Input Stand Saturation value

and Press ENTER

+80.0nA

25.0℃

100%

1

Содержание innoCon 6800O

Страница 1: ...WALL MOUNTING JENSPRIMA INSTRUMENTS Online Dissolved Oxygen Analyzer Operating Instruction innoCon 6800O PRIMA_O2_3BC6051 NORMAL ATC REL WASH 4 88mg l C 25 0 1 2 O O2...

Страница 2: ......

Страница 3: ...r Manual www jensprima com Contents Safety 2 3 5 5 6 7 8 9 10 11 12 13 13 16 17 Display Maintenance General 6 5 Analysis System 14 14 14 15 Zero Calibration Concentration Calibration Output Interface...

Страница 4: ...factured by jensprima Damages to the instrument caused by the use of sensors not manufactured by jensprima within the warranty period will not be covered by the warranty Only technicians authorized by...

Страница 5: ...ion methods standard 4 20mA RS485 output and a variety of control function output Both Chinese and English menus users can do programming on their own Sewage andindustrial wastewater treatment sh farm...

Страница 6: ...r Temperature load 600 RS485 Modbus RTU IP65 138 x 138mm 144 x 144 x 110mm innoSens 420 DO Electrode Suitable for measuring dissolved oxygen in 0 20 00mg L ppm water such as sewage treatment aquacultu...

Страница 7: ...138mm on the instrument case put the instrument into the case from the front of the panel of the case insert screws from the back cover and then lock the fastening block to the fastening bar from the...

Страница 8: ...25 Note the electrodes innoCon 6800 Monitor Manual Page 6 PRIMA_pH_1BC6051 NORMAL ATC REL WASH 7 00 pH C 25 0 1 2 O innoSens 425 0 20 00ppm 0 200 0ppb Immersion type O ring O ring Molded shell electro...

Страница 9: ...ith special training Notes Power supply must be cut o otherwise the instrument and the sensor are likely to be damaged Page 7 power 90 260Vac sensor live wire of power supply neutral wire of power sup...

Страница 10: ...atus of the washing relay ATC MTC manual Menu key Back key Enter key Arrows key Return to previous menu Accept the gure input or menu chosen or o the backlight of the display Entry into the setup mode...

Страница 11: ...2 4 00mA 1 AnalysisSystem Temperature System Measurement System Pressure Salinity Damping 1 Output Analog current 1 Analog current 2 Relay1 Relay2 Relay3 1 General Language Serial Number Restore 1 Int...

Страница 12: ...previous menu 2 Select Restore and press Enter 3 Press upward and downward arrows to select YES or NO and press Enter General 2 Select Language press Enter and you will get a list of available languag...

Страница 13: ...Enter to save the setting and then reture to previous menu Measurement System Analysis System Page 11 Temperature System Temp Unit Pt 1000 0 0 025 0 Sensor Type Temp Offset Manual Temp 1 Measurement...

Страница 14: ...n reture to previous menu Analog Current2 4 00mA 20 00mA 100 0 0 00mA 000 0 Optional range 10 0 110 0 14 0 230 0 Optional range 0 0 120 0 32 0 248 Optional range 1 00 1 00mA mA Offset 1 Relay 1 1 Ente...

Страница 15: ...y 1 Relay 2 Press Enter to test whether Relay3 is closed or open Press Enter to output 4mA 12mA 20mA compulsorily Press Enter to output 4mA 12mA 20mA compulsorily Press Enter to test whether Relay2 is...

Страница 16: ...tore 1 Zero Calibration Current Current nA value of the electrode Enter the current standard temperature setting range available 0 60 0 C Current nA value of the electrode Current temperature of stand...

Страница 17: ...nd Saturation value and Press ENTER 80 0nA 25 0 100 1 1 Wash the electrode dry and put it into saturated air input the current saturated air temperature the next interface Saturation Calibration Curre...

Страница 18: ...is used after a few months storage replace the electrolyte rst before using If the membrane cap fails eg the response time becomes longer the value deviationbecomese larger or the calibration does not...

Страница 19: ...0xD5CA 1 byte 1 byte 2 byte 2 byte 2 byte Order Order of data data data Data content CRC16 Page 17 1 RS485 The client computer responds in the following format When the instrument displays 02 it means...

Страница 20: ...Jensprima Instruments Shanghai Co Ltd Web www jensprima com Email info jensprima com 2017 Jensprima all rights reserved Informa on contained in this document is subject to change without no ce...

Отзывы: