5

USING VISIONPLUS pH630FA

A. Power On/Off

Press the

“On/Off ”

key to turn the unit on. If the unit is running

then you can press the

“On/Off”

key to turn the unit off. The unit

will automatically turn off after 30 minutes of no key activity.

B. Select Buffer

1.

Make sure unit is in “Measure Mode”.

2.

Press the

“Cal”

key for 5 seconds, the unit will display

7.00 to indicate you successfully changed from

“6.86/4.00/9.18” set to “7.00/4.01/10.01” set.

3.

Press the

“Cal”

key for 5 seconds again, the unit will

display 6.86 to indicate you successfully changed from

“7.00/4.01/10.01” set to “6.86/4.00/9.18” set.

[

Note:

There is no need to repeat this procedure every

time unless one decides to change the buffer settings.]

C. Calibrate pH

The user can select one, two or three point pH calibration.

1.

Rinse the electrode & ATC assembly in distilled water

and immerse them in the first buffer solution. The

temperature displayed is the buffer temperature.

2.

Press

“Cal”

key to initiate calibration, the buffer icon will

be on, the “HOLD” icon will flash until the unit detects a

stable reading, When a stable reading is reached, the

“HOLD” icon will be on. Press the

“Enter”

key to save

the calibration, wait a second, the first point has been

calibrated and the unit is ready to be sloped at the

second buffer.
[

Note:

At this moment, Press the

“Hold”

key, the unit

will exit the calibration mode.

Single point

calibration is

complete.]

3.

Remove the electrode & ATC assembly from the first

buffer. Rinse them in distilled water and immerse them in

the second buffer solution. The unit will display the

temperature of the second buffer.

4.

The first buffer icons will be on, the second and third

buffer icons will flash and the “HOLD ” icon will flash

until the unit detects a stable reading, When a stable

Содержание VisionPlus pH630FA

Страница 1: ...Operation Manual MODEL pH630FA Microcomputer Based pH Temperature Pocket Meter pH630FA...

Страница 2: ...and Meter 2 VISIONPLUS pH630FA OVERVIEW 3 A Meter Description 3 B LCD Display 3 OPERATION MODES AND KEYPAD OPERATIONS 4 A Operation Modes 4 B Keypad Operations 4 USING VISIONPLUS pH630FA 5 A Power on...

Страница 3: ...can operate within specifications for approximately 2 to 3 hours after low battery indicator appears 1 Shown in the right figure 2 Take off the battery cover 3 Remove all of the old batteries and inse...

Страница 4: ...ATC assembly Electrode cap is not shown B LCD Display 1 Calibration buffer s and number of calibration point 2 Low battery indicator 3 Calibration mode indicator 4 HOLD mode indicator 5 pH reading 6...

Страница 5: ...either be 7 00 7 00 4 01 10 01 pH or 6 86 6 86 4 00 9 18 pH B Keypad Operations Key Operation Mode Duration Function Hold Measure 0 second Holds current measurement reading Press again to resume measu...

Страница 6: ...Rinse the electrode ATC assembly in distilled water and immerse them in the first buffer solution The temperature displayed is the buffer temperature 2 Press Cal key to initiate calibration the buffer...

Страница 7: ...hird buffer icons will flash and the HOLD icon will flash until the unit detects a stable reading When a stable reading is reached the HOLD icon will be on Press the Enter key to save the calibration...

Страница 8: ...e the old electrode from the electrode collar 3 Insert a new electrode make sure the electrode fit back into the meter correctly 4 Screw back the electrode collar 5 Soak the electrode in a pH 4 buffer...

Страница 9: ...or 14 00 pH ovr Over Range or Recalibrate ovr 32 0 to 140 F pH CAL a pH OFFSET b pH SLOPE a Offset 7 00 pH mV 100 mV Offset Offset 6 86 pH mV 108 3 mV b New slope ideal slope by 30 Use a new buffer so...

Страница 10: ...n pH Buffer Temperature range pH Electrode Offset recognition AUTO 14 2 F to 212 F 32 0 F to 140 F 100 mV at 7 00 pH 108 3 mV 91 7 mV at 6 86 pH pH Electrode Slope recognition 30 at 4 00 4 01 9 18 and...

Страница 11: ...e repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives before returning items for any reason When applying for authorization have the model and ser...

Отзывы: