JENCO VisionPlus pH630FA Скачать руководство пользователя страница 4

3

VISIONPLUS pH630FA OVERVIEW

A. Meter Description

1.

Battery cap

2.

LCD screen

3.

Keypad

4.

Electrode collar

5.

Electrode & ATC assembly

(Electrode cap is not shown.)

B. LCD Display

1.

Calibration buffer(s) and number of calibration point

2.

Low battery indicator

3.

Calibration mode indicator

4.

HOLD mode indicator

5.

pH reading

6.

Temperature reading

Hold

Cal

Enter

1

2

3

4

5

PT-70



Germany

pH

HOLD

CAL

F

6.86 4.00 9.18 7.00 4.01 10.01

1

2

3

4

5

6

Содержание VisionPlus pH630FA

Страница 1: ...Operation Manual MODEL pH630FA Microcomputer Based pH Temperature Pocket Meter pH630FA...

Страница 2: ...and Meter 2 VISIONPLUS pH630FA OVERVIEW 3 A Meter Description 3 B LCD Display 3 OPERATION MODES AND KEYPAD OPERATIONS 4 A Operation Modes 4 B Keypad Operations 4 USING VISIONPLUS pH630FA 5 A Power on...

Страница 3: ...can operate within specifications for approximately 2 to 3 hours after low battery indicator appears 1 Shown in the right figure 2 Take off the battery cover 3 Remove all of the old batteries and inse...

Страница 4: ...ATC assembly Electrode cap is not shown B LCD Display 1 Calibration buffer s and number of calibration point 2 Low battery indicator 3 Calibration mode indicator 4 HOLD mode indicator 5 pH reading 6...

Страница 5: ...either be 7 00 7 00 4 01 10 01 pH or 6 86 6 86 4 00 9 18 pH B Keypad Operations Key Operation Mode Duration Function Hold Measure 0 second Holds current measurement reading Press again to resume measu...

Страница 6: ...Rinse the electrode ATC assembly in distilled water and immerse them in the first buffer solution The temperature displayed is the buffer temperature 2 Press Cal key to initiate calibration the buffer...

Страница 7: ...hird buffer icons will flash and the HOLD icon will flash until the unit detects a stable reading When a stable reading is reached the HOLD icon will be on Press the Enter key to save the calibration...

Страница 8: ...e the old electrode from the electrode collar 3 Insert a new electrode make sure the electrode fit back into the meter correctly 4 Screw back the electrode collar 5 Soak the electrode in a pH 4 buffer...

Страница 9: ...or 14 00 pH ovr Over Range or Recalibrate ovr 32 0 to 140 F pH CAL a pH OFFSET b pH SLOPE a Offset 7 00 pH mV 100 mV Offset Offset 6 86 pH mV 108 3 mV b New slope ideal slope by 30 Use a new buffer so...

Страница 10: ...n pH Buffer Temperature range pH Electrode Offset recognition AUTO 14 2 F to 212 F 32 0 F to 140 F 100 mV at 7 00 pH 108 3 mV 91 7 mV at 6 86 pH pH Electrode Slope recognition 30 at 4 00 4 01 9 18 and...

Страница 11: ...e repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives before returning items for any reason When applying for authorization have the model and ser...

Отзывы: