background image

 

10 

2. 

Press 

“Mode” key until “pH” icon and “AUTO” icon lit up.   

 
3. 

Rinse the pH electrode and ATC/temp probe with distilled 
water and immerse in the sample to be measured. Remove 
any air bubbles trapped around the probe by shaking or 
stirring the probe.   

 
4. 

Press the 

“Mea.” key. The “WAIT” icon will start flashing. The 

unit is waiting for a stable reading. The display will track the pH 
value as sensed by the pH electrode and the ATC/Temp 
probe.   

 
5. 

When the 

“WAIT” icon disappears, the reading is then “LOCK” 

and will not respond to further changes from the sample. The 
pH value shown is the pH value of the sample at the displayed 
sample temperature. 

 

[

Note: 

For samples that are inherently unstable, the unit will 

not AUTOLOCK. In this case, use the pH NON- AUTOLOCK 
mode for measurements.] 
 

b. 

Measurement with manual temperature compensation in 
the pH-AUTOLOCK mode. 

 
1. 

Connect the pH electrode to the BNC connector of the unit. 
The 

“MAN” icon will lit up. Set unit to display the sample 

temperature by pressing the up and down keys(0.0 to 100.0

). 

 
2. 

Repeat steps 2~5 of 

“Measurement with an ATC/Temp 

probe in the pH- AUTOLOCK mode

”.

 

 

c. 

Measurement with an ATC/Temp probe in the pH NON- 
AUTOLOCK mode. 

 
1. 

Connect the pH electrode to the BNC connector and the 
ATC/Temp probe to the ATC/Temp connector of the unit. The     
“ATC” icon will lit up. 

 
2. 

Press 

“Mode” key until “pH” icon lit up.   

 
3. 

Rinse the pH electrode and ATC/temp probe with distilled 
water and immerse in the sample to be measured.   

 
4. 

Allow sufficient time for the display to stabilize. The instrument 
will display the pH value of the sample at the displayed sample 

 

temperature. 

 
 
 
 
 

Содержание VisionPlus pH6175

Страница 1: ...0 Operation Manual MODEL pH6175 Microcomputer Based pH ORP Temperature Benchtop Meter pH6175 ...

Страница 2: ...de 8 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 9 C pH Measurements 9 a Measurement with an ATC Temp probe in the pH AUTOLOCK mode 9 b Measurement with manual temperature compensation in the pH AUTOLOCK mode 10 c Measurement with an ATC Temp probe in the pH NON AUTOLOCK mode 10 d Measurement with manual temperature compensation in the pH NON AUTOLOCK mode 11 D T...

Страница 3: ...easurements enables the unit to automatically sense the end point and LOCK the display to indicate the end point value of a measurement AUTOLOCK and user prompts help eliminate most errors in determining pH and mV values resulting in precise repeatable and error free measurements The pH6175 can also be used in non AUTOLOCK mode The model pH6175 is available with pH ORP and ATC Automatic Temperatur...

Страница 4: ...t the batteries into the meter follow the procedure outlined below 1 Use two hands to flip the two buckles and battery cover to expose the battery compartment Figure 1 2 Note the polarity and insert the six AAA batteries into the battery compartment correctly 3 Replace the battery cover 4 Figure 1 Battery compartment DISPLAY KEYS FUNCTIONS A Display 1 2 3 4 5 6 7 9 10 12 11 8 Figure2 Active LCD sc...

Страница 5: ...efficiency value is less over than 75 5 ATC MAN ATC indicator will be displayed if a temperature probe is connected otherwise the MAN indicator will be displayed 11 MAIN DISPLAY For pH mV and probe efficiency values 6 CAL This will be displayed when the unit enters into the calibration mode 12 SECONDARY DISPLAY For temperature in display B Keys Stand M ode Clear On Off Press and hold this key for ...

Страница 6: ...used to manually enter the temperature values They have no effect on the unit when operating in ATC mode M ea Eff Stand Slope M ode Clear M ea Eff Stand Slope M ode Clear Stand Slope The Stand and Slope keys are used for pH calibration of the unit Pressing and holding the Stand key while turning on the power will change the buffer set M ea Eff Stand Slope M ode Clear Mea Eff The key is used to bri...

Страница 7: ... temperature reading to stabilize then press and hold Stand key for 5 seconds to calibrate The WAIT icon will flash until the unit detects a stable reading Once the unit calibrates the first point the selected buffer remains lit up while the remaining two buffers start to flash The unit is ready to be sloped at the second buffer Note If the first buffer solution is 7 00 or 6 86 at this moment pres...

Страница 8: ...ation values stored in internal memory 2 Connect the pH electrode to the BNC connector of the unit MAN icon will lit up pH icon and AUTO icon will lit up One of the buffer in the pre selected buffer set will start to flash 3 Rinse the pH probes in distilled water and immerse it in the first buffer solution Adjust the temperature reading to that of the first buffer using the up or down keys 0 0 to ...

Страница 9: ...oint calibration is complete If the first buffer solution is 4 00 4 01 9 18 or 10 00 at this moment the unit will automatically exit the calibration mode Single point calibration is complete 4 Rinse the pH and ATC Temp probe in distilled water and immerse them in the second buffer solution either 4 00 4 01 or 9 18 10 01 Allow temperature reading to stabilize then press Slope key to calibrate The u...

Страница 10: ...ins lit up while the remaining two buffers start to flash The unit is ready to be sloped at the second buffer Note If the first buffer solution is 7 00 or 6 86 at this moment Press the Mode key the unit will exit the calibration mode Single point calibration is complete If the first buffer solution is 4 00 4 01 9 18 or 10 00 at this moment the unit will immediately exit the calibration mode Single...

Страница 11: ...ot AUTOLOCK In this case use the pH NON AUTOLOCK mode for measurements b Measurement with manual temperature compensation in the pH AUTOLOCK mode 1 Connect the pH electrode to the BNC connector of the unit The MAN icon will lit up Set unit to display the sample temperature by pressing the up and down keys 0 0 to 100 0 2 Repeat steps 2 5 of Measurement with an ATC Temp probe in the pH AUTOLOCK mode...

Страница 12: ...RP electrode to the BNC connector of the unit 2 Press Mode key until mV icon and AUTO icon lit up 3 Rinse electrode with distilled water and immerse it in sample to be measured 4 Press the Mea key The WAIT icon will start flashing The unit is waiting for a stable reading The display will track the mV value as sensed by the ORP electrode 5 When the WAIT icon disappears the reading is then LOCK and ...

Страница 13: ... coefficient of the pH buffers 4 00 6 86 9 18 4 01 7 00 10 01 0 4 01 6 98 9 46 4 01 7 11 10 32 5 4 00 6 95 9 39 4 01 7 08 10 25 10 4 00 6 92 9 33 4 00 7 06 10 18 15 4 00 6 90 9 28 4 00 7 03 10 12 20 4 00 6 88 9 23 4 00 7 01 10 06 25 4 00 6 86 9 18 4 01 7 00 10 01 30 4 01 6 85 9 14 4 01 6 98 9 97 35 4 02 6 84 9 10 4 02 6 98 9 93 40 4 03 6 84 9 07 4 03 6 97 9 89 45 4 04 6 83 9 04 4 04 6 97 9 86 50 4...

Страница 14: ...he electrode and ATC Temp probe settled to within 30 of the buffer value 2 Buffer 4 00 4 01 9 18 and 10 01 is not correct 3 pH electrode slope is off by more than 30 of ideal slope 1 Allow sufficient time for the electrode and ATC Temp probe to stabilize re press Slope key to continue the calibration procedure 2 Check if the correct buffer is used 3 Replace the buffer and or the pH electrode Press...

Страница 15: ...o 60 0 C pH Electrode Offset recognition 100 mV at pH 7 00 or 108 3 91 7mV at pH6 86 pH Electrode Slope recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input impedance 1012 Ω Temperature sensor Thermistor 10 kΩ at 25 C Power Six AAA Batteries Calibration Back up EEPROM Audio Feedback All Touch Keys End Point Sensing Hold Yes Display pH mV Temp 22mm 14 5mm high LCD Ambient Temperature Range 0 to 50 C...

Страница 16: ...zation have the model and serial number handy including data regarding the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss Jenco will not be responsible for damage resulting from careless or insufficient packing A fee will be charged on all authorized returns NOTE Jenco reserves the right to make improve...

Отзывы: