6
4.
CAL-
This will be displayed when the
unit enters into the calibration
mode.
10.
MAIN DISPLAY-
For Conductivity, Salinity and
TDS values.
5.
AUTO-
Auto ranging indicator
11.
SECONDARY DISPLAY-
For temperature in
℃
.
6.
ATC/MAN-
ATC indicator will be displayed
if a temperature probe is
connected otherwise the MAN
indicator will be displayed.
12.
ppt-
Parts per thousand, indicates
Salinity measurement.
B. Keys
On/Off-
Press and hold this key for 2 seconds to power on
and shut off the meter. 0 second to Turn Bluetooth on
or off
M
ea.
E
ff
.
S
tand
S
lope
M
ode
C
lear
Mode-
Selects display mode. Pressing this key changes the
display sequentially to display pH-AUTO, mV
(RmV)-AUTO, pH , mV (RmV) , Conductivity ,
Salinity and TDS . The calibration values will not be
affected by changing the display modes. In
“calibration”,
press
“Mode”
key to exit calibration
mode.
M
ea.
E
ff
.
S
tand
S
lope
M
ode
C
lear
Clear-
It is used to clear the unit when error signal appears. It
clears all calibration values stored in the internal
memory.
Under normal use the key will not be activated unless
pressed and held for 5 seconds to prevent accidental
erasing stored memory.
In pH, mV (Rmv) or conductivity (salinity, TDS) mode,
when the
“Clear”
key is pressed for 5 seconds, all
segments of the LCD will be lit on.
After 2 seconds the
unit will enter the pH-AUTO mode, mV-AUTO mode
or conductivity mode, In the pH mode, the “AUTO”
and “CAL” icons will be displayed and one of the
buffer values in the pre-selected buffer set will begin
to flash. This means that the unit is un-calibrated and
requires new pH calibration before use. In the ORP
mode, the “AUTO” and “mV” icons will be displayed.
This means that the unit is un-calibrated and requires
new ORP calibration before use. In the conductivity
mode, the “AUTO” icon will be displayed. This means
that the unit is requires new conductivity calibration
before use.
M
ea.
E
ff
.
S
tand
S
lope
M
ode
C
lear