17

3. Standard solution of 111.9 mS/cm at 25

: Accurately weight

out 74.264 grams of research grade dried Potassium Chloride

(KCL). Dissolve in 1000 ml of distilled water.

[Note:

You can store the unused portion of the standard

solution in a plastic container for six months but the air space

between the cap and the solution must be kept to an absolute

minimum. Storing the excess solution below 4

can increase

the storage life. If you have any doubt of the accuracy of the

stored solution, a fresh batch should be prepared.

]

BLUETOOTH WIRELESS TRANSMISSION FUNCTION

Description

The 6377B is a multi-parameter pH / ORP / Conductivity /

Salinity / TDS temperature Bluetooth benchtop meter that pairs

with iOS and Android tablets to enable the same user

experience as users’ favorite apps while providing graphing,

guided calibration procedures, GLP-compliant data storage,

data sharing, cloud data storage and user-defined compliance

policies.

a. Prepare 6377B for Bluetooth connection.

Press the

“On/Off”

key for 2 second to turn on the 6377B

device, then press and release the On/Off button again to turn

on 6377B Bluetooth connection. The Bluetooth light on the

6377B shall fast blink and it is ready for Bluetooth connection.

b. Pair the 6377B to your tablet.

Tap on the “Settings” tab at the lower right of the app

screen.

Tap on the “Pair” button.

The tablet shall search any nearby JENCO Bluetooth

devices within close proximity. When the app discovers

and displays your 6377B, tap the “Connect” button.

Once the 6377B is connected to the tablet, the device

Bluetooth light shall blink slowly.

After the app has been installed, calibrate your 6377B to

ensure the best reading quality.

Troubleshoot

The app is running and the 6377B is on, but for some

reason they won't connect.

If the tablet’s Bluetooth is turned on, and the app continues to

display the reconnecting indicator message, close the app and

Содержание VisionPlus 6377B

Страница 1: ...Operation Manual MODEL 6377B Microcomputer Based pH ORP Cond TDS Salt Temperature Benchtop Meter 6377B...

Страница 2: ...nts 11 D Temperature Measurements 12 E mV Offset 12 F mV RmV Measurements 13 G Conductivity Calibration 13 H Conductivity Salinity TDS Measurements 15 I pH Buffers 16 J Preparing Conductivity Standard...

Страница 3: ...taneously along with user prompts and mode indicators The unit prompts the user through calibration and measurement procedures An AUTOLOCK feature for both pH and mV measurements enables the unit to a...

Страница 4: ...E BATTERIES To insert the batteries into the meter follow the procedure outlined below 1 Use a Philip screw driver and unlock the battery cover by turning the screw driver in the counter clockwise dir...

Страница 5: ...ORP Display 1 2 3 4 5 6 7 9 10 12 11 8 Figure 3 Active LCD screen 1 WAIT This will be displayed when the unit is still waiting for a stable reading or end point sensing 7 AUTO AUTOLOCK mode indicator...

Страница 6: ...e is connected otherwise the MAN indicator will be displayed 11 MAIN DISPLAY For pH mV and probe efficiency values 6 CAL This will be displayed when the unit enters into the calibration mode 12 SECOND...

Страница 7: ...tand Slope Mode Clear Clear It is used to clear the unit when error signal appears It clears all calibration values stored in the internal memory Under normal use the key will not be activated unless...

Страница 8: ...ue 3 In the ORP mode press and hold this key for 5 seconds the LCD will display the electrode Offset value 4 In the Conductivity Salinity or TDS mode press and hold this key for 5 seconds the LCD will...

Страница 9: ...automatically exit the calibration mode Single point calibration is complete 4 Rinse the pH and ATC Temp probe in distilled water and immerse them in the second buffer solution either 4 00 4 01 pH or...

Страница 10: ...the first buffer solution is 1 68 4 00 4 01 9 18 10 01 or 12 46 pH at this moment the unit will automatically exit the calibration mode Single point calibration is complete 5 Repeat steps 4 of Calibr...

Страница 11: ...ration 6 The unit calculates and compensates for the pH electrode slope deviation corresponding to the values of the calibrated buffers After calibration press and hold Mea Eff key for about 5 seconds...

Страница 12: ...ple to be measured Remove any air bubbles trapped around the probe by shaking or stirring the probe 4 Press the Mea Eff key The WAIT icon will start flashing The unit is waiting for a stable reading T...

Страница 13: ...20 0 2 Repeat steps 2 to 4 of Measurement with an ATC Temp probe in the pH NON AUTOLOCK mode D Temperature Measure The 6377B can measure temperature independently with the ATC temp probe without using...

Страница 14: ...TOLOCK In this case use the mV RmV NON AUTOLOCK mode for measurements b Measurement in the mV RmV NON AUTOLOCK mode 1 Connect the optional combination ORP electrode to the BNC connector of the unit 2...

Страница 15: ...calculate temperature compensated conductivity The default value is 25 To change the temperature reference press the Up and Down keys to adjust the value between 15 to 25 Press Slope Enter key to save...

Страница 16: ...probe constant to a known value by pressing the Up and Down keys You can now input the K value from 70 to 130 of the probe basic cell constant Press the Slope Enter key to exit calibration and return...

Страница 17: ...rately weight out 0 746 grams of research grade dried Potassium Chloride KCL Dissolve in 1000 ml of distilled water 2 Standard solution of 12 90 mS cm at 25 Accurately weight out 7 4365 grams of resea...

Страница 18: ...liant data storage data sharing cloud data storage and user defined compliance policies a Prepare 6377B for Bluetooth connection Press the On Off key for 2 second to turn on the 6377B device then pres...

Страница 19: ...This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference...

Страница 20: ...r value 2 Buffer 1 68 4 00 4 01 9 18 10 01 and 12 46 pH is not correct 3 pH electrode slope is off by more than 30 of ideal slope 1 Allow sufficient time for the electrode and ATC Temp probe to stabil...

Страница 21: ...ective 3 Incorrect K constant value input 1 Sample cannot be tested 2 Decontaminate clean cell or replace cell 3 Input correct K value over during calibration 10 0 to 120 0 1 Incorrect standard soluti...

Страница 22: ...ure 10 0 to 120 0 C 0 1 C 0 2 C pH buffer recognition 1 68 7 00 4 01 10 01 12 46 pH or 1 68 6 86 4 00 9 18 12 46 pH pH Temperature compensation AUTO MAN 10 0 C to 120 0 C pH Buffer Temperature range 0...

Страница 23: ...Service Department alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be...

Страница 24: ...Jenco Electronics Ltd 6F NO 81 Sec 2 Chang an E Rd Jhongshan District Taipei City 104 Taiwan TEL 886 2 2508 2928 FAX 886 2 2508 2938 E Mail sales jenco com tw Website www jenco com tw Shanghai Jenco I...

Отзывы: