In the “Delete Each” mode, use

“Up”

and

“Down”

key to

select data to be deleted. Then press

“Enter”

key to delete.

Deletion is now complete. The next set of saved data will

automatically move up a slot in the storage site.

5.

Press

“Mode”

key to exit “Delete” mode.

SENSOR MAINTENANCE

Sensor caps last longer if stored in a moist environment and kept

clean and free of scratches in the paint layer. Erratic readings can

result from damaged or dirty caps. If unstable readings occur from

large scratches in the paint or dye layer, replace the sensor cap.

See the Maintenance and Troubleshooting section of this manual for

information on how to clean or replace the sensor cap. The average

sensor cap replacement interval is 12 to 18 months, although they

may last longer.

Avoid substances that may damage probe materials such as

concentrated acid, caustics and strong solvents. Probe materials

include urethane (cable), Xenoy (probe body), acrylic (sensor cap)

and acetal and stainless steel (probe guard).

To prevent the sensor cap from drying out, store the probe in the

gray calibration/storage sleeve with a moistened, clean sponge. In

the event the sensor cap dries out, rehydrate it following the

instructions in the Maintenance and Troubleshooting section of this

manual.

A. Cleaning the Sensor Cap

The Sensor Cap should be kept clean since some types of fouling

can consume or produce oxygen or block the water sample from

reaching the sensor which could affect the dissolved oxygen

measurements.

To clean the Sensor Cap, gently wipe away any fouling with a lens

cleaning tissue that has been moistened with water. Do not use

organic solvents to clean the Sensor Cap. Using an organic solvent

to clean the Sensor Cap may cause permanent damage to the cap.

For example, alcohol will dissolve the outer paint layer and other

organic solvents will likely dissolve the dye in the cap. Dissolving or

scratching the paint or dye layer will cause erratic dissolved oxygen

measurements.

To disinfect the probe, soak the probe in 3% hydrogen peroxide for

15 to 30 minutes and then rinse it with clean water.

Содержание 9031M

Страница 1: ...Operation Manual MODEL 9031M Basis Portable DO Temperature Meter 9031M...

Страница 2: ...6 OPERATIONAL PROCEDURES 8 A Entering Sensor Cap Coefficients into the 9031M 8 B Dissolved Oxygen Calibration 9 C Dissolved Oxygen Measurements 10 D Sampling Modes continuous or manual 11 E Save Reca...

Страница 3: ...031M measures dissolved oxygen in water using optical lifetime luminescence technology and uses a digital signal to send information between the instrument and probe Key advantages of the 9031M includ...

Страница 4: ...ure outlined below Figure 1 Battery compartment 1 Use a screw driver to remove the four screws and battery cover to expose the battery compartment Figure 1 2 Replace the 9 V battery 3 Replace the batt...

Страница 5: ...tor 6 K1 K2 K3 K4 K5 KC Sensor cap coefficient indicators when entering new sensor cap coefficient 7 Temperature unit 8 Temperature values 9 mg L dissolved oxygen unit 10 Full Recall Each All Delete I...

Страница 6: ...Delete mode press these keys to select between the Delete Each and Delete All mode In Delete Each mode view to be deleted data and data site numbers by pressing these keys CAL In Measurement mode pre...

Страница 7: ...OBE A Probe Throughout the manual the term probe refers to the end of the cable where the sensor is located the term sensor refers to the Light Sensing Dissolved Oxygen sensing portion of the cable pr...

Страница 8: ...term storage 3 Reinstall the probe guard by sliding it carefully over the sensor and then threading it onto the cable probe assembly with a clockwise rotation Figure 5 4 Locate the gray calibration s...

Страница 9: ...essing the ENTER key to confirm each entry 5 After confirming the last value for K1 the instrument will prompt you for K2 Follow the instructions in step 3 and 4 to enter and confirm the value for K2...

Страница 10: ...bration mode The CAL icon appears on the LCD The main screen will display 1013 mBar factory default value The meter is now ready for atmospheric pressure calibration 4 To change the pressure factor us...

Страница 11: ...ange the pressure factor use the Up and Down keys to adjust the value between 500 and 1125 mBar Press Enter key to save the new value and the unit will automatically go into the next calibration param...

Страница 12: ...e Recall and Delete Data a Saving readings to memory 1 In Measurement mode press the Enter key to save data The Save icon with the corresponding site number will lit up for a brief moment to indicate...

Страница 13: ...p and acetal and stainless steel probe guard To prevent the sensor cap from drying out store the probe in the gray calibration storage sleeve with a moistened clean sponge In the event the sensor cap...

Страница 14: ...sponge or immersed in water see Sensor Storage for more information If the sensor dries out refer to the Rehydration procedure in this manual Refer to Figure 7 below when following the instructions f...

Страница 15: ...Cap must remain in a moist environment see Sensor Storage for storage recommendations If the sensor cap is allowed to dry out it is likely to drift slightly at the beginning of your next study unless...

Страница 16: ...ibration If erroneous readings persist inspect the sensor cap for damage cracks large scratches in paint or dye layer replace if necessary enter new coefficients and then attempt to recalibrate the in...

Страница 17: ...tabilization time for dissolved oxygen and temperature 3 Check sensor cap Clean rehydrate or replace as necessary 4 Contact JENCO Tech Support DO readings are inaccurate 1 Verify temperature readings...

Страница 18: ...o 1125 mBar Salinity compensation 0 0 to 40 0 ppt Calibration Up to 2 point Calibration Back up EEPROM Datalogging capabilities 50 data sets Automatic shut off function 10 minutes of non use Audio Fee...

Страница 19: ...product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives befor...

Страница 20: ...om Jenco Electronics Ltd 6F NO 81 Sec 2 Chang an E Rd Jhongshan District Taipei City 104 Taiwan TEL 886 2 2508 2928 FAX 886 2 2508 2938 E Mail sales jenco com tw Website www jenco com tw Shanghai Jenc...

Отзывы: