JENCO 9031M Скачать руководство пользователя страница 11

10

switch to “Measurement” mode. Calibration is now complete. If

“Mode”

key is pressed instead of the

“Enter”

key, any

changes made will be cancelled and the previous calibration

settings will be retained and will automatically switch to

“Measurement” mode.

b.

2-point calibration

1. Connect the LSDO probe to the unit and turn the unit on.

2. Place the sensor in a solution of zero DO.( A zero DO solution

can be made by dissolving approximately 8 ~ 10 grams of

sodium sulfite (Na2SO3) into 500 mL tap water or DI water.

Mix the solution thoroughly. It may take the solution 60 minutes

to be oxygen-free.).Press

“Mode”

key to select “Dissolved

Oxygen in %” mode.

3. Wait 3 to 5 minutes for the dissolved oxygen and temperature

reading to stabilize, press the

“CAL”

key to enter the

“Calibration” mode. The “CAL” icon appears on the LCD. The

main screen will display “1013 mBar” (factory default value).

The meter is now ready for atmospheric pressure calibration.

4. To change the pressure factor, use the

“Up”

and

“Down”

keys

to adjust the value between 500 and 1125 mBar. Press

“Enter”

key to save the new value and the unit will

automatically go into the next calibration parameter which is

the value of 100% saturation in air. If

“Mode”

key is pressed

instead of the

“Enter”

key, any changes made will be

cancelled and the previous calibration settings will be retained.

5. In this interface, the user can view the calibration value flashing

in the secondary display. Press

"Up"

or

"Down"

key to select

the display to "0.0". Once the value in the main display

stabilizes, press

Enter

key to save the new value and the

unit will automatically go into the second point calibration. If

“Mode”

key is pressed instead of the

“Enter”

key, any

changes made will be cancelled and the previous calibration

settings will be retained.

6.

Repeat steps 5 to 6 of

“1-point DO% water-saturated air

calibration”

for second point calibration.

C. Dissolved Oxygen Measurements

Press

“Mode”

key to choose the dissolved oxygen in % mode or

dissolved oxygen in mg/L mode. Rinse the LSDO probe with

distilled water and immerse it in the sample to be measured.

Содержание 9031M

Страница 1: ...Operation Manual MODEL 9031M Basis Portable DO Temperature Meter 9031M...

Страница 2: ...6 OPERATIONAL PROCEDURES 8 A Entering Sensor Cap Coefficients into the 9031M 8 B Dissolved Oxygen Calibration 9 C Dissolved Oxygen Measurements 10 D Sampling Modes continuous or manual 11 E Save Reca...

Страница 3: ...031M measures dissolved oxygen in water using optical lifetime luminescence technology and uses a digital signal to send information between the instrument and probe Key advantages of the 9031M includ...

Страница 4: ...ure outlined below Figure 1 Battery compartment 1 Use a screw driver to remove the four screws and battery cover to expose the battery compartment Figure 1 2 Replace the 9 V battery 3 Replace the batt...

Страница 5: ...tor 6 K1 K2 K3 K4 K5 KC Sensor cap coefficient indicators when entering new sensor cap coefficient 7 Temperature unit 8 Temperature values 9 mg L dissolved oxygen unit 10 Full Recall Each All Delete I...

Страница 6: ...Delete mode press these keys to select between the Delete Each and Delete All mode In Delete Each mode view to be deleted data and data site numbers by pressing these keys CAL In Measurement mode pre...

Страница 7: ...OBE A Probe Throughout the manual the term probe refers to the end of the cable where the sensor is located the term sensor refers to the Light Sensing Dissolved Oxygen sensing portion of the cable pr...

Страница 8: ...term storage 3 Reinstall the probe guard by sliding it carefully over the sensor and then threading it onto the cable probe assembly with a clockwise rotation Figure 5 4 Locate the gray calibration s...

Страница 9: ...essing the ENTER key to confirm each entry 5 After confirming the last value for K1 the instrument will prompt you for K2 Follow the instructions in step 3 and 4 to enter and confirm the value for K2...

Страница 10: ...bration mode The CAL icon appears on the LCD The main screen will display 1013 mBar factory default value The meter is now ready for atmospheric pressure calibration 4 To change the pressure factor us...

Страница 11: ...ange the pressure factor use the Up and Down keys to adjust the value between 500 and 1125 mBar Press Enter key to save the new value and the unit will automatically go into the next calibration param...

Страница 12: ...e Recall and Delete Data a Saving readings to memory 1 In Measurement mode press the Enter key to save data The Save icon with the corresponding site number will lit up for a brief moment to indicate...

Страница 13: ...p and acetal and stainless steel probe guard To prevent the sensor cap from drying out store the probe in the gray calibration storage sleeve with a moistened clean sponge In the event the sensor cap...

Страница 14: ...sponge or immersed in water see Sensor Storage for more information If the sensor dries out refer to the Rehydration procedure in this manual Refer to Figure 7 below when following the instructions f...

Страница 15: ...Cap must remain in a moist environment see Sensor Storage for storage recommendations If the sensor cap is allowed to dry out it is likely to drift slightly at the beginning of your next study unless...

Страница 16: ...ibration If erroneous readings persist inspect the sensor cap for damage cracks large scratches in paint or dye layer replace if necessary enter new coefficients and then attempt to recalibrate the in...

Страница 17: ...tabilization time for dissolved oxygen and temperature 3 Check sensor cap Clean rehydrate or replace as necessary 4 Contact JENCO Tech Support DO readings are inaccurate 1 Verify temperature readings...

Страница 18: ...o 1125 mBar Salinity compensation 0 0 to 40 0 ppt Calibration Up to 2 point Calibration Back up EEPROM Datalogging capabilities 50 data sets Automatic shut off function 10 minutes of non use Audio Fee...

Страница 19: ...product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives befor...

Страница 20: ...om Jenco Electronics Ltd 6F NO 81 Sec 2 Chang an E Rd Jhongshan District Taipei City 104 Taiwan TEL 886 2 2508 2928 FAX 886 2 2508 2938 E Mail sales jenco com tw Website www jenco com tw Shanghai Jenc...

Отзывы: