12
Temperature Coefficient
The unit uses the temperature coefficient to calculate
temperature compensated conductivity. The default value is
1.91%. To change the temperature coefficient, use the “UP”
and “DOWN” keys to adjust the value between 0 and 4.00%.
Press “ENTER” to save the new value and the unit will go into
the next calibration parameter automatically. If “MODE” is
pressed instead of “ENTER”, any changes made will be
cancelled and the previous calibration settings will be
retained.
Temperature Reference
The unit uses the temperature reference value to calculate
temperature compensated conductivity. The default value is
25
℃
. To change the temperature coefficient , use the “UP”
and “DOWN” keys to adjust the value between 15
℃
to 25
℃
.
Press “ENTER” key to save the new value and the unit will go
into the next calibration parameter automatically. If “MODE”
key is pressed instead of the “ENTER” key, any changes
made will be cancelled and the previous calibration settings
will be retained.
%
1.91
CAL
25.0
℃
25.0
℃
CAL
25.0
℃