SPECIFICATIONS                             

  

 

Display Range 

Resolution 

Accuracy 

 

pH 

-2.00 to 16.00 pH

0.01 pH

±0.01±1digit 

mV 

-1999 to 1999 mV

1 mV 

±0.1% 

Temperature  -10.0 to 120.0 °C

0.1 °C 

±0.5°C 

 

 

pH buffer recognition

 

pH 7.00, 4.01, 10.01 or pH 

6.86, 4.00, 9.18 

pH Temperature compensation

AUTO/MAN –10.0°C to 

120.0 °C 

pH Buffer Temperature range

 

0°C to 60.0°C 

pH Electrode Offset recognition

±90 mV at pH 7.00 or 6.86 

pH Electrode SLOPE 
recognition

 

±30% at pH 4.00, 4.01, 9.18 

and 10.01 

Input impedance

 >10

12

 

Temperature sensor

 Thermistor, 

10 

k

 at 25°C 

Power

 9Volt 

battery 

Battery Life

 ~1000 

Hours 

Calibration Back-up

 EEPROM 

Automatic shut off function 

30 minutes of non-use

 

Audio Feedback

 

All Touch Keys 

End Point Sensing & Hold

 Yes 

Display (pH / mV : Temp)

 

12mm : 8mm high LCD 

Ambient Temperature Range

 

0 to 50 °C 

Relative Humidity

 

At 90% RH 

Case

 IP65 

waterproof

   

Dimensions (W x D x H)

 

70mm x 198mm x 37mm 

Weight 

 

 

260 grams 
(Batteries included) 

 

 

 

 

 

 

Содержание 6010N

Страница 1: ...Operation Manual Hand held pH ORP Temperature Meter 6010N...

Страница 2: ...OLOCK mode 8 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 9 C pH Measurements 10 a Measurement with an ATC Temp probe in the pH AUTOLOCK mode 10 b Measurement with ma...

Страница 3: ...HOLD the display to indicate the end point value of a measurement AUTOLOCK and user prompts help eliminate most errors in determining pH and mV values resulting in precise repeatable and error free me...

Страница 4: ...NSTALLING THE BATTERIES The 6010N meter is packaged with one 9V battery required for operation To insert the batteries into the meter follow the procedure outlined below Figure 1 Battery compartment 1...

Страница 5: ...he electrode 3 HOLD This will indicate that the reading is frozen during AUTOLOCK mode 8 pH mV Unit and mode indicators 4 ATC MAN ATC indicator will be displayed if a temperature probe is connected ot...

Страница 6: ...it Pressing and holding the STAND key while turning on the power will change the buffer set to the other available buffer set MEA EFF The key is used to bring the unit out of the AUTOLOCK condition wh...

Страница 7: ...and temperature readings with the pH icon on but AUTOLOCK icon off 4 mV mode The unit will display mV and temperature readings with the mV icon on but AUTOLOCK icon off OPERATIONAL PROCEDURES A Buffe...

Страница 8: ...temperature reading to stabilize then press and hold STAND key for 2 seconds to calibrate The WAIT icon will flash until the unit detects a stable reading Once the unit calibrates the first point the...

Страница 9: ...he pH probe in distilled water and immerse it in the second buffer solution either 4 00 4 01 or 9 18 10 01 then press SLOPE key to calibrate The WAIT icon will flash until the unit detects a stable re...

Страница 10: ...en the efficiency value is over than 75 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 1 Turn the unit on Press CLEAR key for 5 seconds all LCD elements will lit up The...

Страница 11: ...and ATC temp probe with distilled water and immerse them in the sample to be measured Remove any air bubbles trapped around the probe by shaking or stirring the probe 4 Press the MEA EFF key The WAIT...

Страница 12: ...the pH NON AUTOLOCK mode for measurements c Measurement with an ATC Temp probe in the pH NON AUTOLOCK mode 1 Connect the pH electrode to the BNC connector and the ATC Temp probe to the ATC Temp connec...

Страница 13: ...Press the MEA EFF key The WAIT icon will start flashing The unit is waiting for a stable reading The meter will track the mV value as sensed by the ORP electrode 5 When the WAIT icon disappears and th...

Страница 14: ...0 4 01 6 98 9 46 4 01 7 11 10 32 5 4 00 6 95 9 39 4 01 7 08 10 25 10 4 00 6 92 9 33 4 00 7 06 10 18 15 4 00 6 90 9 28 4 00 7 03 10 12 20 4 00 6 88 9 23 4 00 7 01 10 06 25 4 00 6 86 9 18 4 01 7 00 10...

Страница 15: ...rrect 3 pH electrode SLOPE is off by more than 30 of ideal SLOPE 1 Allow sufficient time for the electrode and ATC Temp probe to stabilize re press SLOPE key to continue the calibration procedure 2 Ch...

Страница 16: ...ecognition 90 mV at pH 7 00 or 6 86 pH Electrode SLOPE recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input impedance 1012 Temperature sensor Thermistor 10 k at 25 C Power 9Volt battery Battery Life 10...

Страница 17: ...product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives befor...

Страница 18: ...coi com Website www jencoi com Jenco Electronics Inc 1F NO 11 Lane 370 Sec 6 Zhongxiao E Rd Nangang Dist Taipei Taiwan TEL 886 2 2782 3226 FAX 886 2 2782 3234 Shanghai Jenco Instruments Ltd 18 Wang Do...

Отзывы: