ERROR DISPLAYS AND TROUBLESHOOTING       

 

 

 

Main  

Display

 

Possible cause(s) 

Corrective Action(s)

 

"Er1" 

1. “STAND” was pressed   

before the electrode and 
ATC/Temp probe settled 
to /-1.5 pH of the 
buffer value. 

 
 
2. pH electrode offset is 

greater / less than +/-1.5 
pH. 

 
 
3. pH electrode is faulty.   

1. Press “CLEAR” key, allow 

sufficient time for the 
electrode and ATC/Temp 
probe to stabilize, re-press 
“STAND” key to start the 
calibration procedure. 

 
2. Replace the buffer and /or 

the pH electrode. Press 
“CLEAR” key to recalibrate 
meter. 

 
3. Replace electrode. 

"Er2" 

1. “SLOPE” was pressed     

before the electrode and 
ATC/Temp probe settled 
to within 30% of the 
buffer value. 

 
2. Buffer 4.00, 4.01, 9.18 

and 10.01 is not correct.

 
3. pH electrode SLOPE is 

off by more than 30% of 
ideal SLOPE. 

 

1. Allow sufficient time for the 

electrode and ATC/Temp 
probe to stabilize, re-press 
“SLOPE” key to continue 
the calibration procedure. 

 
2. Check if the correct buffer is 

used. 

 
3. Replace the buffer and /or 

the pH electrode. Press 
“CLEAR” key to recalibrate 
meter. 

"Er3" 

1. Temperature is out of the

0.0 to 60.0°C range. 

 1. Bring the buffer temperature 

within range. 

"over" 

/ ”undr” 

1. Measured pH is out of 

the –2.00 to 16.00 pH 
range. 
 

2. Measured mV is out of 

the –1999 to 1999 mV 
range. 

 
3. Measured temperature is 

out of the –10.0 to 
120.0°C range. 

1. Bring sample pH into the 

correct measuring range. 
 

 
2. Bring sample ORP into the 

correct measuring range. 

 
 

3. Bring sample temperature 

into the correct measuring 
range. 

[Note: 

If the meter still does not perform normally after the above 

measures are taken, call 

Jenco 

Service Department.

 

14

Содержание 6010N

Страница 1: ...Operation Manual Hand held pH ORP Temperature Meter 6010N...

Страница 2: ...OLOCK mode 8 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 9 C pH Measurements 10 a Measurement with an ATC Temp probe in the pH AUTOLOCK mode 10 b Measurement with ma...

Страница 3: ...HOLD the display to indicate the end point value of a measurement AUTOLOCK and user prompts help eliminate most errors in determining pH and mV values resulting in precise repeatable and error free me...

Страница 4: ...NSTALLING THE BATTERIES The 6010N meter is packaged with one 9V battery required for operation To insert the batteries into the meter follow the procedure outlined below Figure 1 Battery compartment 1...

Страница 5: ...he electrode 3 HOLD This will indicate that the reading is frozen during AUTOLOCK mode 8 pH mV Unit and mode indicators 4 ATC MAN ATC indicator will be displayed if a temperature probe is connected ot...

Страница 6: ...it Pressing and holding the STAND key while turning on the power will change the buffer set to the other available buffer set MEA EFF The key is used to bring the unit out of the AUTOLOCK condition wh...

Страница 7: ...and temperature readings with the pH icon on but AUTOLOCK icon off 4 mV mode The unit will display mV and temperature readings with the mV icon on but AUTOLOCK icon off OPERATIONAL PROCEDURES A Buffe...

Страница 8: ...temperature reading to stabilize then press and hold STAND key for 2 seconds to calibrate The WAIT icon will flash until the unit detects a stable reading Once the unit calibrates the first point the...

Страница 9: ...he pH probe in distilled water and immerse it in the second buffer solution either 4 00 4 01 or 9 18 10 01 then press SLOPE key to calibrate The WAIT icon will flash until the unit detects a stable re...

Страница 10: ...en the efficiency value is over than 75 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 1 Turn the unit on Press CLEAR key for 5 seconds all LCD elements will lit up The...

Страница 11: ...and ATC temp probe with distilled water and immerse them in the sample to be measured Remove any air bubbles trapped around the probe by shaking or stirring the probe 4 Press the MEA EFF key The WAIT...

Страница 12: ...the pH NON AUTOLOCK mode for measurements c Measurement with an ATC Temp probe in the pH NON AUTOLOCK mode 1 Connect the pH electrode to the BNC connector and the ATC Temp probe to the ATC Temp connec...

Страница 13: ...Press the MEA EFF key The WAIT icon will start flashing The unit is waiting for a stable reading The meter will track the mV value as sensed by the ORP electrode 5 When the WAIT icon disappears and th...

Страница 14: ...0 4 01 6 98 9 46 4 01 7 11 10 32 5 4 00 6 95 9 39 4 01 7 08 10 25 10 4 00 6 92 9 33 4 00 7 06 10 18 15 4 00 6 90 9 28 4 00 7 03 10 12 20 4 00 6 88 9 23 4 00 7 01 10 06 25 4 00 6 86 9 18 4 01 7 00 10...

Страница 15: ...rrect 3 pH electrode SLOPE is off by more than 30 of ideal SLOPE 1 Allow sufficient time for the electrode and ATC Temp probe to stabilize re press SLOPE key to continue the calibration procedure 2 Ch...

Страница 16: ...ecognition 90 mV at pH 7 00 or 6 86 pH Electrode SLOPE recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input impedance 1012 Temperature sensor Thermistor 10 k at 25 C Power 9Volt battery Battery Life 10...

Страница 17: ...product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives befor...

Страница 18: ...coi com Website www jencoi com Jenco Electronics Inc 1F NO 11 Lane 370 Sec 6 Zhongxiao E Rd Nangang Dist Taipei Taiwan TEL 886 2 2782 3226 FAX 886 2 2782 3234 Shanghai Jenco Instruments Ltd 18 Wang Do...

Отзывы: