Page 6
Note:
If light level is selected prior to selecting a calibration/sensitivity factor, an error
message Set Factor will appear in place of light level readings (see 15Q for selecting
factors). If OD and % Trans, are selected and no baseline 100% value has been set and
error Set 100% will appear see 3G8 below to learn more about setting 100%)
3G3. Integrate:
(Requires Display Units set to light level) Summation of data, records exposure/dosage
(i.e. W/cm2 becomes J/cm2 in integrate mode).
3G4. Min-Max:
Holds the display at the present value until a new minimum (lower reading) or new
maximum (higher reading) is measured, then updates the display accordingly.
Note:
Min-Max also serves as a Peak FW. See section 15F Min/Max (and FW Peak)
3G5. Record:
Starts and stops (toggles between) continuous recording of readings and creates a new
file with time, instantaneous, min, max and average measurements.
3G6. Zero:
Takes a measurement and creates a zero value by subtracting the zero measurement
reading from all future readings (see 3F2 above to learn more about assuring an
accurate zero).
3G7. Hold
:
Freezes the numerical display value on screen.
3G8. 100%:
Takes a measurement and saves it as the 100% baseline for Transmission and OD.
3G9. Capture:
Takes a single measurement. Captured measurements are appended to the single
sample file.
3G10. Quick Start Care and Handling:
i.
Always store the meter in the carrying case when not in use and only ship in a
carrying case with protective packing to prevent shock and damage. No items
should be placed above or below the meter inside the carrying case.
ii.
Do not expose meter to direct sunlight or high heat sources as the internal
temperature can rapidly increase and cause erroneous readings and damage.
iii.
Standard operating temperature is rated at 0
Deg C
to 40 Deg C.
iv.
Do not apply any force to the display.
v.
Do not apply pressure to extended stand.
vi.
Opening the housing voids warranty. Housing is not water-tight; avoid moisture,
liquids and chemicals.