Page 17
15N. Dim Display After time settings
:
Sets the time in minutes before the display will dim to save on battery life. Times range
from 5 min - 60 min and also includes Never.
15O. Turn Off Display After time settings
:
Sets the time in minutes before the display will shut off to save on battery life. Times
range from 5 min - 60 min and also includes Never.
Maintenance
15P. Set Date and Time:
Allows the user to set the date and time which are then used to create file names and
date/time stamps on the saved .csv files as discuss in 16 above.
15Q. Manage Calibration Factors >
:
Creating calibration factors and use of customer generated factors can cause
errors in light level readings. The calibration factors listed in the Optical
Calibration Certificates included are N.I.S.T. traceable and are performed in our
ISO17025 accredited lab by trained and experienced technicians. Use of altered
calibration factors can void traceability.
15Q.1. Factor Number:
Numbers range from 1-20. The JEL-2400 stores up to 20 calibration factors which
are selected using Factor number as shown in section in 15B.
15Q.2. Factor Description:
Enter a nick name, component model&Serial number, or sensor stack i.e. lux sensor,
UVC#234
.
(Note: spaces are converted to a “-”).
15Q.3. Factor:
Units Enter the units for the readout i.e. lux (Note: Sp
aces are converted to a “-”).
15Q.4. Sensitivity Factor:
Enter sensitivity factor from the calibration certificate. 2.09e-09.
15Q.5. Resulting Light Level:
Allows the user to enter a known value for the light level under test and create a
calibration factor to force the meter to read the correct value.
i.e. LED with 40 lumens is measured. Enter 40, press set. A new calibration factor
will be created to read 40 and auto-range up and down through other light
intensities.
15Q.6. Factor Adjustment:
User can enter a value to alter the calibration factor by for example if adding a
neutral density filter with attenuation of 10, enter 10 and press set. A new calibration
factor, that has changed by 10, will be entered in the next available channel.