S-II & J-II Chairs

Operating / Service Manual

 

25

TROUBLESHOOTING

Problem

Solution

Possible Cause

Chair is difficult to rotate.

Chair will not lower unless weight is 

applied.

Chair base moves or feels unstable.

All chair bases must be bolted securely 

to the floor to prevent tipping.

Chair  base  is  not  bolted  securely  to 

the floor.

Chair slowly loses height after weight 

is applied.

 

Chair feels "spongy" when entering.

Air  bubbles  may  have  entered  the 

hydraulic  system.    This  may  occur 

from lifting on chair top without having 

rotation lock applied.

Bleed air from hydraulic system.  

(see "Maintenance" section)

Chair will not lift at all (electric base).

Test power outlet with another device 

to assure power is available.

Chair  base  is  not  plugged  into  a 

powered outlet.

Chair will not lift to full height.

Add hydraulic fluid to hydraulic system. 

(contact JEDMED for assistance)

Hydraulic system is low on fluid.

Backrest requires assistance to 

return to upright position.

Adjust or replace gas-spring cylinder 

on bottom of chair.  

(contact JEDMED for assistance)

Gas-spring  cylinder  is  weak  or  mis-

adjusted.

Brake will not lock chair rotation.

Re-adjust rotation lock mechanism.

(contact JEDMED for assistance)

Rotation  lock  mechanism  mis-

adjusted.

Rotation lock mechanism mis-

adjusted.

Re-adjust rotation lock mechanism.

(contact JEDMED for assistance)

Entire lift piston comes out of chair 

base when chair top is lifted.

Lock chair rotation (& height) by firmly 

lifting on pump lever / footpedals

Rotation lock mechanism is not applied.  

Must be applied when lifting or moving 

chair.

Chair lift piston is noisy / squeals.

Lightly lubricate lift piston with hydraulic 

oil or grease.

Adjust rotation lock mechanism 

(contact JEDMED for assistance)

Tighten entire rotation lock mechanism 

(contact JEDMED for assistance)

Re-position base shroud.

Lift piston requires lubrication.

Rotation lock mechanism is too tight.

(squeal)

Entire rotation lock mechanism is loose. 

(chatter)

Base shroud (plastic cover) is touching 

lift piston.

Rotation lock is ON

Release  rotation  lock  by  depressing 

pump lever / footpedals.

Содержание J-II

Страница 1: ...Operating Service Manual An Employee Owned Company S II J II Chairs IMPORTANT Read all instructions before assembling or using this device...

Страница 2: ...we serve Thank you for your support and we hope that your experience with this product will be a rewarding one which will result in a lasting relationship with JEDMED Your suggestions on how we can im...

Страница 3: ...tification Operation 8 13 Cleaning Disinfection 14 Maintenance 15 Dimensions 16 17 Reorder Items 18 Specifications 19 Section Page CONTENTS OPERATION SERVICE OPERATING Troubleshooting 24 Parts Diagram...

Страница 4: ...ECAUTIONS KEY ALERTS user to the possibility of PERSONAL INJURY WARNING The following conventions are used throughout this manual ALERTS user to the possibility of DAMAGE to the EQUIPMENT CAUTION ALER...

Страница 5: ...r out into a table J II chairs offer manual backrest recline only with a fixed position footrest Both chairs include synchronized armrests which lay out flat as backrest is reclined When entering or e...

Страница 6: ...o skid Slowly remove carton Carefully cut stretch wrap away from chair Use extreme care not to damage the chair upholstery or finish Preferably use scissors in lieu of razor knife Unbolt chair base fr...

Страница 7: ...y is required other than mounting base to the floor ASSEMBLY Options are normally supplied fully assembled and installed at the factory If options are ordered after initial purchase follow instruction...

Страница 8: ...N Headrest Backrest Armrest Seat Footrest Lift base A B C D E F Main Parts Headrest height adjustment handwheel Headrest tilt adjustment lever Recline lever 1 2 3 Chair Controls J II Chair A B C D E F...

Страница 9: ...e Lever Place hand behind backrest cushion to support patient in chair Using other hand push down on Recline Lever Move chair backrest to desired recline position Release Recline Lever to lock backres...

Страница 10: ...ng holes 1 2 3 4 Lift Base Controls Electric Base Pump bar Raise Lower Rotation lock Floor mounting holes 1 2 Lift Base Controls Manual Base All lift bases must be bolted to floor to protect against a...

Страница 11: ...se by repeatedly pumping downward on pump bar When pumping pump each downward stroke only to the first feel of resistance Pumping past the first stop point will open the valve to lower the chair Lower...

Страница 12: ...airs Operating Service Manual 12 Light Installed OPTIONS Booster In Place Examination Light Pediatric Booster Seat Headrests Exam Light Kit Standard Headrest Double articulating Headrest IDENTIFICATIO...

Страница 13: ...S II J II Chairs Operating Service Manual 13 Blank...

Страница 14: ...ient contact areas should be wiped down with a cold disinfecting solution and a disposable cloth Do not allow solution to remain on product any longer than is necessary to ensure sufficient disinfecti...

Страница 15: ...k base rotation Locking base rotation also locks the lift piston from moving If chair is lifted when hydraulic piston is not locked air may be introduced into the hydraulic lines If so the chair may s...

Страница 16: ...torized Lift Base 110V A 20 to 28 B 23 00 C 41 50 04 1547 JEDMED S II Chair with Motorized Lift Base 220V A 20 to 28 B 23 00 C 41 50 76 25 to 80 50 21 50 13 00 9 50 23 50 8 75 to 13 00 20 00 20 25 41...

Страница 17: ...5 to 13 00 23 50 22 50 6 00 60 75 to 65 00 20 20 25 04 1550 JEDMED J II Chair with Manual Pump Base A 19 5 to 27 B 26 75 C 41 25 04 1555 JEDMED J II Chair with Motorized Lift Base 110V A 20 to 28 B 23...

Страница 18: ...ative or JEDMED Customer Service at Item Bulb for exam light Replacement Parts 99 4160 Catalog 5416 JEDMED Court St Louis MO 63129 314 845 3770 phone 314 845 3771 fax e mail info jedmed com website ww...

Страница 19: ...al Order 100 4 7 50 60 0 2kW Special Order Lifting Capacity 350 lbs Base Diameter 23 inches Lifting Stroke 20 28 5 inches from floor Manual Lift Base Motorized Lift Base Manual Hydraulic Base Lifting...

Страница 20: ...S II J II Chairs Operating Service Manual 20 Blank...

Страница 21: ...SERVICE INFORMATION...

Страница 22: ...Blank...

Страница 23: ...ce Manual 23 SERVICE INFORMATION ALWAYS unplug unit from main power outlet before cleaning disinfecting maintenance or service WARNING Service should only be performed by a qualified technician Contac...

Страница 24: ...eturn to upright position Adjust or replace gas spring cylinder on bottom of chair contact JEDMED for assistance Gas spring cylinder is weak or mis adjusted Brake will not lock chair rotation Re adjus...

Страница 25: ...MBER ALL VERSIONS 23 1 1 LAB 113 LABEL UNDER PENALTY OF LAW 22 1 1 LAB 019 LABEL JEDMED 21 2 2 KA04010 LEVER GRIP 20 3 3 KA01006 HOLE PLUG 0 375 NICKEL PLATED STEEL 19 4 4 K119027 LOCKNUT M8 1 25MM TH...

Страница 26: ...ECURELY 26 1 1 502 AW001 LABEL SERIAL NUMBER ALL VERSIONS 25 1 1 LAB 113 LABEL UNDER PENALTY OF LAW 24 1 1 LAB 019 LABEL JEDMED 23 2 2 KA04010 LEVER GRIP 22 3 3 KA01006 HOLE PLUG 0 375 NICKEL PLATED S...

Страница 27: ...for pedal 114 Nut lock adjusting 115 Lock Mechanism cast iron 116 Lock Lever 117 O Ring for shaft 118 Pump Shaft 119 Connection Bar 120 Up Pedal right 121 Down Pedal left 122 Control Lever right 123...

Страница 28: ...5587 Bottom Socket Tube HLP 4004H Socket Tube Hold Down Washer HLP 5585A Socket Tube Hold Down Washer Screw HLP 5586A 3 Stem HLP 5506H See BOM for Variations Spacer Tube Brite Chrome HLP 5503HC See B...

Страница 29: ...NOTES...

Страница 30: ...5416 JEDMED Court St Louis MO 63129 314 845 3770 phone 314 845 3771 fax e mail info jedmed com website www JEDMED com...

Отзывы: