4.2.10 Duty
Range
Accuracy
0.1%~99%
1%
4.2.11 capacitance
Range
Resolution
Accuracy
4nF 0.001nF
40nF 0.01nF
400nF 0.1nF
4
µ
F 1nF
40
µ
F 10nF
±(
3% rdg + 3dgt
)
200
µ
F 100nf
Range
Accuracy
0.1Hz~10M
Hz
0.1Hz~10MHz
5. Measurement operation
5-1 DC & AC voltage measurement
Warning
To avoid harms to you or damage to the meter from electric
shock. Please do not attempt to measure voltage higher than
DC/AC 1000V although readings may be obtained.
The DC voltage range are 400.0mV, 4.000V, 40.00V , 400.0V and 600V and
then. The AC voltage ranges are 4.000V, 40.00V, 400.0V and 600V.
To measure DC or AC voltage:
1.
Insert the red test lead into the “
V
Ω
” input terminal and the black test
lead into the COM terminal.
2.
Set the rotary switch to
DC or AC
range.
3.
Connect the test lead across with the object under testing.
The measured value will be show on the LCD display.
Note:
When DC or AC voltage measurement has been completed,
disconnect the connection
between the testing lead and the circuit under testing.
5-2. Resistance measurement
The resistance range are: 400.0
Ω
, 4.000K
Ω
,40.00K
Ω
, 400.0K
Ω
,
4.000M
Ω
.
40.00M
Ω
.
To measure resistance, connect the meter as follows:
1.
Insert the red test lead into the ”
V
Ω
” terminal and the black test lead into
the COM terminal.
2.
Set the rotary switch to proper resistance range..
3.
Connect the test lead across with the object under testing.
The measured value will be show on the LCD display.
Note:
The test lead can add 0.1
Ω
to 0.2
Ω
of error to resistance