background image

 

6

 

 

Thank you for purchasing the ECHO-FT. 

 

Intended use 

This home blood pressure monitoring device is indicated for use in the noninvasive 

measurement of systolic and diastolic blood pressure and heart  rate in adult age 

18  and  over,  in  a  home  care  environment.  ECHO-FT  is  designed  to  enable  a 

person who suspects some trouble with blood pressure to check it at home. 

 

Fuzzy logic and Oscillometric method   

Your  unit  employs  fuzzy  logic  method  to  inflate  and  an  oscillometric  for 

measurement of blood pressure and pulse, which refers to check  blood pressure 

by measuring oscillation in the cuff while the artery of the arm oscillate reflecting 

contracts and relaxes every heart beat.   

 

Data transmissible and analyzing program 

This device is data transmissible. You can transmit readings saved in the memory 

of the device to your personal computer, and then mail them to your physician to 

get  advice  from  her/him  at  home.  ECHO-FT  is  a  time  saving  device  partially 

replacing  a  necessity  of  your  visit  to  your  physician.  Additionally  the  data 

transmitted  to  your  PC  can  be  analyzed  and  tracked  sophisticatedly  with   

analyzing software optionally offered by the manufacturer.   

 

Software ‘Hour 24’ 

‘Hour24’ is a software for windows. It takes data transmitted from ECHO-FT and 

enable to send them to the doctor. 

Additionally the data can be analyzed and tracked with it. 

 

Noise detecting 

As  ECHO-FT  is  contrived  to  detect  moving  or  noising  during  measurement 

process and displays it both on the real time and recalled readings on LCD. You 

or your physician can determine if the readings are reliable.   

 

Voice message   

It can make you monitor blood pressure easily and know the result clearly without 

                                                             

INTRODUCTION 

Содержание ECHO-FT

Страница 1: ...LCD indicator 14 PRECAUTIONS FOR MEASUREMENT 15 BATTERY INSERTION REPLACEMENT 18 1 How to insert the batteries 18 2 Replacement of the batteries 18 CONNECTING OF AC ADAPTER 19 SETTING THE DATE AND TIME 19 1 Entering the mode to set date and time 19 2 Setting up 19 SOUND ON OFF AND VOLUME CONTROL 20 DATA TRANSMISSION 21 MEASUREMENT 23 1 Wrapping the cuff 23 2 Correct measuring posture 24 3 Measurin...

Страница 2: ...his manual is made for informational purpose and this manual and product are not meant to be a substitute for the advice provided by your own physician or other medical professional You should not use this information contained in this manual or this equipment for diagnosis or treatment of health problem or prescription of medication If you have or suspect that you have a medical problem consult y...

Страница 3: ...is the pressure creating the heart pulls blood back through it Blood pressure is recorded as systolic over diastolic as shown below systole occurs during diastole occurs during the contraction and the relaxation and empting of the heart filling of the heart 2 What makes my blood pressure fluctuate Blood pressure changes in response to many factors These factors include changing in the temperature ...

Страница 4: ...eadings to give to your physician to determine and diagnose and also provides you the information when your blood pressure falls or rises to enable to manage your health between doctor visits 5 What is the benefit of data transmissible blood pressure monitor Without going to a hospital you can get physician s advice with data transmissible blood pressure monitor ECHO FT is a Home to Hospital bi di...

Страница 5: ...tyle than those inherent in a clinical environment It let you learn when your blood pressure falls or rises Also it let your physician optimize the treatment program corresponding your circadian rhythm of blood pressure Therefore it is important to take regular daily monitoring and keep clear records ECHO FT is a time and cost saving healthcare networking instrument ...

Страница 6: ...nsmissible You can transmit readings saved in the memory of the device to your personal computer and then mail them to your physician to get advice from her him at home ECHO FT is a time saving device partially replacing a necessity of your visit to your physician Additionally the data transmitted to your PC can be analyzed and tracked sophisticatedly with analyzing software optionally offered by ...

Страница 7: ...ined by a trained observer using the cuff stethoscope auscultation method within the limits prescribed by the American National Standard Electronic or Automated Sphygmomanometers Note Blood pressure constantly fluctuates during of a day so that monitoring should be done at least three times and determined Occasional high or low readings do not mean your monitor is malfunctioning or you are ill The...

Страница 8: ...manual to denote information of special emphasis Property damage refers to a wide range of damage involving your house household goods domestic animals and pets etc CAUTION Important information to indicate the presence of a hazard in which may arise severe personal injury death of substantial property damage if the warning is ignored NOTE Important information to notify the user of operation or m...

Страница 9: ...tions Please keep this manual together with the appliance since it is compiled to provide the maximal convenience to those who use this device Defective unit or accessories that is shipped off from customer to Jawon medical Co or its authorized dealer for service must be packed in the safe way and the replacement carton Shipping and insurance cost for returning of defective unit must be prepaid by...

Страница 10: ...ion Level B for Noise immunity by standard of IEC60601 1 2 Electromagnetic Compatibility Requirements This blood pressure meter complies with the EN1060 1 1995 Non invasive Sphygmomanometers general requirements as well as EN1060 3 1997 supplementary requirements for electro mechanical blood pressure measuring system CLASSIFICATION COMPLIANCE WITH STANDARD NOTE 1 Jawon s accessory equipment connec...

Страница 11: ...11 ...

Страница 12: ...for storage Where the ambient temperature deviates 10 40 for operation Where the atmospheric pressure deviates 70kPa 700mbar 106kPa 1060mbar Where the humidity is above 95 for storage and 75 for operation Where the unit is exposed to spray or splashing water Where the unit is exposed to dust Where the unit is exposed to water vapor Where the unit is exposed to high density oil vapor Where the unit...

Страница 13: ...r Do not insert the batteries with their polarities in wrong direction The International Electro technical Commission IEC has established a set of symbols for medical electrical equipment that classify a connection or warn of any potential hazards The symbols are shown below Save these instructions for your safety Isolated patient connection IEC 60601 1 Type BF I and O on power switch represent ON...

Страница 14: ...e mode Memory view sound off and volume control button Button to transmit data Button to enter to the mode to set date and time Number up button to set date and time USB cable To use data transmission refer to owners manual of software Hour 24 how to use or how to install PART NAMES FUNCTIONS ...

Страница 15: ...sting Displays during deflation DATE 10 01 TIME 12 00 month day hour minute Starts flash when purse is detected Displays when the batteries need to be changed Memory indicator Displays when sound is ON Indicates noise or movement sensed 01 Indicates the number of stored results ...

Страница 16: ...tion Lack of sleep or constipation causes the blood pressure to rise Cold weather causes the blood pressure to go up Maintain indoor temperature at 20 or so and warm your body before checking Drinking caffeinated beverages such as coffee tea cokes and also smoking can cause your blood pressure to rise Moving or speaking while checking causes errors PREPARATION FOR MEASUREMENT ...

Страница 17: ...the monitoring arm To fit the cuff on the naked arm is advised When you roll up your sleeve ensure it does not constrict the blood flow in your arm The position sitting legs crossed or using on a lower table than heart level may increase abdominal pressure that resulted in elevated blood pressure When an over pressurization rises or to stop the pressurizing press O I button The cuff will stop infl...

Страница 18: ...f the chair and desk to enable your monitoring arm to place at the same level as your heart When monitoring arm is higher than your heart level the blood pressure will be measured lower When monitoring arm is lower than your heart level the blood pressure will be measured higher PREPARATION FOR MEASUREMENT ...

Страница 19: ...compartment as shown above 1 If the polarities of the battery are positioned incorrectly the unit does not work When you change the batteries use the same type Remove the used batteries immediately to protect the product from damage When you are not going to use the unit for an extended period of time approximately three months or more remove the batteries Note If the readings are not transmitted ...

Страница 20: ...ll appear on the screen and 00 flashes 2 Setting up Press SET UP button to set the year Whenever SETUP button is pressed or you keep it pressing the number will increase gradually or rapidly by 01 Its range is 00 to 99 After setting the year press CLOCK button again The month indicator the first two numbers in upper LCD line will flash Set it by pressing SET UP button Its range is 01 to 12 SETTING...

Страница 21: ...ound ON OFF If you press I O button around 3 seconds Snd On OFF signal appears on LCD By pressing button to change it Press I O button to complete it 2 Volume control If you keep I O button pressing after appearing Snd signal the signal is changed to voL 1 Select the volume size by pressing button Its range is 1 to 4 Press I O button to complete it Note When you press I O button after setting the ...

Страница 22: ...to enable your physician to optimize the treatment program and determine how much medication is required to you corresponding your circadian rhythm of blood pressure To install an USB driver in a computer follow the guidance contained in manual of software Hour 24 Turn on the power of two appliances after connecting ECHO FT to the computer running Hour 24 USB cable should be used the one supplied ...

Страница 23: ...hs and right upper side when a data to send A window to get input if click a E mail button opens Input a mail address and title and click a Send E mail button A save situation of a file is marked in E mail title and a transmitted figure file and blood pressure measurement file are saved in a computer of a user Hour24 manual reference Ncte Please transmit the data file which use a SAVE function of ...

Страница 24: ...ff hose over your brachial forearm artery and between your middle and ring fingers when it is extended Position the lower edge of the cuff is 1 1 inches above the elbow joint as shown in the picture Strain the cuff with 2 fingers space between the cuff inner liner and the arm Place your arm at the same level as the heart MEASUREMENT Note Take off thick coat sweater or tight fit clothes from the mo...

Страница 25: ...feet flat on the floor and take deep breaths 5 to 6 times and relax 3 Measuring method Press the I O button Wrap the cuff around your arm and press I O button to start There will be a momentary full segment display Automatic pressurization The LCD panel changes to the momentary screen displayed all symbol in LCD edge and 0 in main LCD Note Sit comfortably on a chair with your feet flat on the floo...

Страница 26: ...sensed during deflation symbol appears and flashes When noise is detected during measurement appears on its place on the screen In case that you move or speak is accompanied with a voice warning as Don t move or speak please But monitoring will be continued after a pause When Error occurs due to insufficient inflation re pressurization will be held automatically with a voice massage as Can t measu...

Страница 27: ...with monitoring date and time respectively The reading detected noise or movement during measurement accompanies 1 Storing the readings The readings are stored automatically when the monitoring is completed When its memory capacity is fully occupied the last data replaces the oldest one 2 How to retrieve the stored data Press the button The latest reading will be displayed on main Note Even if you...

Страница 28: ...he data will be displayed respectively in reverse order from the latest one to oldest one with memory number increasing Note 1 Data saved in memory can be all deleted exchange battery or remove an AC adapter 2 Please insert battery in unit when use an AC adapter 3 To completely eliminate saved data take out battery and please press one time of I O button 4 The memory capacity is subject to change ...

Страница 29: ...ust direct sunlight humidity and extreme changes in temperature If this unit is not being used for a long period of time please remove the batteries Do not drop the unit Do not disassemble the main body or the cuff MAINTENANCE AND STORAGE ...

Страница 30: ...ater in outer case dry it immediately with a soft lint free cloth Do not use abrasive or corrosive cleaning agent thinner alcohol benzene or similar solvents which might damage the device It is recommended to clean the unit monthly 2 Cuff Do not wash or dip the cuff in the water Note Do not immerse the unit in water The device exposed to foreign material and exceptional environment need to be chec...

Страница 31: ...he power off then on and monitor again Err 3 When monitoring is carried out the wrong way When a read data is not output as a result of noise or movement detected Please read wrapping the cuff and correct measuring posture in Measurement part page 18 19 Monitor again following the way instructed Err 4 When moving or speaking is sensed during monitoring Please do not move or speak during monitoring...

Страница 32: ...ring monitoring Make sure that your arm position at the same level as your heart then monitor again Re pressurization In case of insufficient inflation The movement of the arm or body during inflation is sensed In case of excessive hypertension or a feeble pulse it is possible to occur re pressurization over 2 times Relax for a few minutes then monitor it again with correct posture Do not move dur...

Страница 33: ... exclusion or limitation of incidental or consequential damages or limitations on how long an implied warranty lasts so the above exclusions or limitations may not apply to you This warranty gives you specific legal rights and you may also have other rights which vary from state to state Should your product prove defective during this warranty period please return the product in the original carto...

Страница 34: ...tments improper maintenance alteration maladjustment of controls modification power surges commercial use of the product or service by anyone other than an authorized service center or acts beyond Manufacturer s control 3 Where applying for solving problem Should you be dissatisfied with the service you received from your retailer please write JAWON MEDICAL CO LTD Seoul office Tel 82 2 587 4056 Fa...

Страница 35: ...0 Humidity 30 75 RH non condensing Storage Environment Temperature 20 to 60 Humidity less than 95 RH non condensing Dimension 102 0 W X 133 0 D X 63 2 H mm Weight Approx 500g Including cuff batteries Pressure Mode Fuzzy logic Automatic inflation Specifications are subject to change without notice Life of battery can be reduced according to environment This manual cannot do whole or a part reproduc...

Страница 36: ...36 ...

Отзывы: