30
(4) Slide the id of the light source unit to remove it (see Fig.5.4).
Note: Do not loosen screws other than the adjustment screw for the stopper.
(5) Turn the deuterium lamp stopper adjustment screw (Fig. 5.5 or Fig. 5.6) and search for a
position where the value displayed on the monitor is a maximum.
(6) Replace the lid of the light source unit.
5.3.3 Consumable Parts
Table5.1 gives consumable parts necessary for maintenance. On any order, please quote the
part names and also the part numbers.
Table 5.1 Consumable parts
Part name
Part number
Description
Halogen lamp (WI)
5330-0099
Deuterium lamp (D
2
)
5330-0094B
Fuse (2.0 A Time-lag fuse)
5840-H102A
5 fuses/set, V-630, 100 V
Fuse (2.5 A Time-lag fuse)
5840-H103A
5 fuses/set, V-650/660/670, 100 V
Fuse (1.6 A Time-lag fuse)
5840-H106A
5 fuses/set, V-630/650/660/670, 200 V