5.4 Pump head check valve assembly replacement
5.4.1 Preparations for check valve assembly replacement
Replace the solvent inside the pump with one of the following solvents (See Section
4.2):
y
Methanol
y
Water
Since this operation involves the disassembly of the pump, solvent or liquid
remaining inside the pump will unavoidably come into contact with fingers and hands.
In addition, a leak may develop during the pressure test. Therefore, a solvent that is
relatively harmless to both humans and the instrument must be selected for use
during check valve assembly replacement. If organic solvents are currently being
used, replace these solvents with methanol. If a buffer solution or other solution
containing salts is being used, replace this solution with water.
WARNING
: Chloroform and acetonitrile are hazardous to the human body and
should not be used for the pressure test.
CAUTION: Since, salts in buffers may precipitate, causing poor instrument
performance, do not use buffers for the pressure test.
Note: Since water tends to leave air bubbles in the pump head, be sure that any
water used for the pressure test has been sufficiently degassed.