GT2
Calibration Guide
Janam Technologies LLC
2
Step 3
Compare the average values from Steps 1 and 2. If these two values are different, the temperature
reading must be calibrated. To prepare for calibration, complete the following steps:
1.
Connect to Wi-Fi, record the IP address on the screen
2.
Input the IP address plus +port 8090 in the PC Chrome window (Example: 192.168.8.2
∶
8090)
3.
Access the control panel by entering the password “123456”
4. Access
S
ettings
by clicking the gear icon on the blue menu bar at the bottom of the menu
Step 4
In the temperature measurement calibration field, enter the difference between the FDA-approved
thermometer (Step 1) and GT2’s temperature reading functionality (Step 2). For the example used,
the value that should be entered is “0.1”.
•
Positive calibration (“compensation”) is between 0.0 ~ 10.0
•
Negative calibration (“compensation”) is between -0.0 ~ -10.0