CM-200GE/CM-200GE-RA/CB-200GE/CB-200GE-RA
- 47 -
In the case of Sequential EPS, (Trigger source is Software)
For each sequence,
The following table shows the minimum trigger interval in asynchronous accumulation mode. In
case of sequential mode, only asynchronous mode is functional. Therefore, the trigger timing
should be set so that the timing is not in synchronous mode.
Full Scan
2/3 Partial
1/2 Partial
1/4 Partial
1/8 Partial
1/2 V Binning
1255 L
895 L
717 L
453 L
323 L
631 L
•
The condition for this table is that shutter speed should be set the same for all sequences.
If the shutter speed is different, the difference of exposure time should be added. It is
recommended to set the exposure time in the order from the shortest to the longer one.
•
For proper operation, It is not allowed to issue a trigger pulse immediately after the
sequence is reset. The trigger must be issued at least 500ms after the end of the sequence.