ADwin System with Ethernet Interface
ADwin
56
ADwin
Installation, manual version 2.1, December 2005
AC
LI
ST / BU
BU
ST
LI / ST
9.2.2 The DIL switches
ADwin-Pro
only
The settings of the DIL switches are part of the specified features of
ADwin-Pro
systems only and can there be changed by the user.
For all other
ADwin
systems please refer to Jäger Computergesteuerte
Messtechnik GmbH or the corresponding representative in your country.
For Pro modules there are 2 types of Ethernet interfaces, where the DIL swit-
ches are set differently.
You can differentiate between the 2 types by their firmware version:
– Up to firmware version 2.5: ENET-1
– From firmware version 2.6 onwards: ENET-2
To do a version check start the program
<ADethflash>
(from Windows start
menu:
Programs
ADwin
). Choose the appropriate module and the option
"
Display
the
status
…
" and confirm with "
Next
". The next window shows
the software version of the module.
ENET-1
Now change the DIL switches. (For the allocations and functions, see below).
The change of the DIL switches will only become active at the next power-up
of the
ADwin
Ethernet interface.
Place the module with the front plate to the left side, so that the handle is closer
to you.
This description refers to the DIL switch block at the lower right of the
ADwin
Ethernet interface printed circuit board (only with the older Pro module
T9+ENET: the DIL switch block at the upper right must not be changed).
Abb. 1 – DIP switch numbers ENET-1
Mode of operation
LED display
Ethernet Interface is
sending
LED
AC
is lighted.
Physical connection
recognized
LED
LI
is lighted.
Restart (the Ethernet
interface)
LED
ST
is off. LED
BU
is lighted for approx. 5 seconds
and then is switched off.
The LED
ST
is now switched on and off quickly, i.e.
you see a more glowing light. Now the Ethernet inter-
face is ready for operation.
Access
The LED
BU
is lighted as long as the
ADwin
system is
accessed via the Ethernet interface. In parallel the
LED
ST
changes its blinking rhythm.
Configuration
If a configuration is successfully accepted, the LED
ST
starts blinking several times before the module
executes a restart with the new configuration.
Ready for operation
(no communication)
LEDs
LI
and
ST
are lighted.
DIL no.
Description
Meaning
Default setting
1
Reprogramming the firm ware
ON = active
OFF
2
Bootloader write-protected
ON = yes
OFF
3
Bootloader deactivated
ON = yes
OFF
4
Configuration blocked
ON = yes
OFF
5…8
No function
–
OFF