JWI-700D/700S 7
→
→
→
Step 3: Install the Indicator (4) on the bracket,
with the bracket aligning with the bracket slot
(8) of the indicator.
Step 4: After connecting load cell connector to load cell port, the installation is completed.
Note: Use the knob pole (7) to adjust the inclination angle of indicator and the screw (6) to adjust
direction of the indicator. After adjusting the indicator to an optimal position, lock the screw.
5. Setting Mode
5-1 Maximum Weighing Capacity & Division Setting
1. Press and hold key
UNIT
and
M-/SET
while powering on the scale. When the window
displays
“
CAP
”
,
release the key and it enters the capacity & division setting
2. Press
MR/
or
TARE/
to shift among
“
CAP
” — capacity setting; “
DIV
”— division setting;
“
accura
” — resolution check.
3. When the window is showing
“
CAP
”, press
M-/SET
to enter capacity setting.
Method1: Press
MR/
or
TARE/
to choose the general used capacity. Press
M+/SAVE
to save and return.