                        JWI-700C                                              

9

       

 

= There will be a warning sound when the material quantity is less than the lower limit,

 

and 

the weight of the material is more than 20 divisions. 

= No sound alarm.

 

8) Press numeric key 

8

 

to set whether to save the   

upper & lower limit of quantity checking.   

.

= Previously set quantity checking values are not retained when the unit is turned on. 

.

= Previously set quantity checking values are retained when the unit is turned on. 

9) Press numeric key 

9

 to set print modes .Options are 

Prt.Pr, Prt.Co and Prt.St. If Prt.Co is chosen as the print   
mode, PC will be automatically selected as the   
external devices. 

.

= manual print 

.

= continuous print 

.

= Stable printing (the weight of the weighted articles should be more than 9 divisions. 

Weighted articles should be removed and the scale goes  back to zero before print out the 
next record.) 

10) Press key 

MC/CK 

to select external devices. 

Options are PC

AX

TP

Godex ,SH

 

ET

TDP and ZEBRA. 

11) Press Key 

MR

 to switch On or OFF RTC 

  function. 

12) When setting is completed, press key 

ZERO

 to save and return to weighing mode. 

  6.

 Calibration Mode 

1) If good linearity of the load cell was not guaranteed, linear calibration is required. After the 
linear calibration is made, there is no need to make single point calibration. 

2) If there is limit in weights, single point calibration is needed .it would be more accurate to 
calibrate the scale with 1/3 or more of the full load. 

3) Here we take 75kg weighing capacity as an example. 

Содержание JWI 700C

Страница 1: ...ion Setting 5 3 Function Setting 7 7 8 6 Calibration Mode 6 1 Linear calibration 6 2 Single point calibration 10 10 7 Operation 7 1 Basic Weighing 7 1 1 Weighing 7 1 2 Tare Preset Tare 7 1 3 Unit Switch Operation 7 2 Basic Counting 7 2 1 Upper Lower Quantity Limit Checking 7 2 2 Entering a Known Unit Weight 7 2 3 Sample Counting ACAI 7 2 4 Accumulation Accumulation Display and Accumulation Clear 1...

Страница 2: ...ich might cause incorrect reading Avoid sudden temperature changes suitable operating temperature is between 5 40 Disconnect the power supply when cleaning the indicator Do not immerse the indicator in water or other liquids 3 Product Introduction 3 1 Specifications Features Specifications Model JWI 700C Input sensitivity 0 13uV D Input voltage range 0 5mV to 16 5mV Load cell excitation DC 5V Up t...

Страница 3: ...ndicator Automatic average weighing unit function Single point and linear calibration available HI OK LO indicator HI OK LO alarm function Full range tare pre tare auto zero tracking Rechargeable battery or AC power Automatic counting accuracy improvement ACAI Accumulation of weight and quantity max 99 pieces Auto power off selection save battery life Suitable for a wide range of bases and load ce...

Страница 4: ...c counting accuracy improvement is functioning 10 M Symbol points at M when scale is in accumulation mode 11 UW Under entering a known unit weight mode Symbol points at UW when entered unit weight is lower than 4 5 of scale division Unit weight is too small for ensuring accurate quantity calculations 12 SMPL Under the sample counting mode Symbol points at SMPL when calculated unit weight is lower ...

Страница 5: ... Clears the indicated input values 7 MC CK key Short press deletes accumulation records long press to initial Upper Lower Quantity Limit setting 8 MR key Recalls and displays the total Accumulation data weight total count and number of weightments and the first 10 records in details 9 UWS key Inputs the indicated unit weight value entered via numeric keys 10 M key Adds the indicated weight or piec...

Страница 6: ...g The charge lamp turns green from red when the recharging is completed which takes about 8 hours Disconnect the scale from power supply when it is fully charged Note Battery is to replaced only by an authorized service dealer Risk of explosion can occur if replaced with the wrong type or connected improperly 4 Installation 4 1Load Cell Connection CONN 1 2 3 4 5 E E S S LOAD CELL Shield LOAD CELL ...

Страница 7: ...cell connector is too big to thread through the bracket separate the bracket by removing the Knob pole 7 see the following pictures Step 3 Install the Indicator 4 on the bracket with the bracket aligning with the bracket slot 8 of the indicator Step 4 After connecting load cell connector to load cell port the installation is completed Note Use the knob pole 7 to adjust the inclination angle of ind...

Страница 8: ...ity and division 5 Press key ZERO to go back to the weighing mode 5 2 Resolution Setting 1 Turn on the power while pressing key SMPL 2 Key in 1123 with the numeric keys 3 Enter the resolution selection mode by pressing key SMPL Numeric Key 1 is to circle through resolutions options Available setting are 1 1 6000 or 1 7500 2 1 15000 and3 1 30000 4 Press to key SMPL save and return to the setting me...

Страница 9: ...ically turns off Options are OFF Non power off 5 10 30 and 60 minutes 5 Press numeric key 4 to set the range in which the Zero indication turns on Options are d0 d1 d2 d3 d4 and d5 d scale division 6 Press numeric key 5 to select serial transmission rates Options are 9600 4800 and 2400 7 Press numeric key 7 to set buzz sounds Options are Un In no Lo and nbEEP There will be a warning sound when the...

Страница 10: ...inuous print Stable printing the weight of the weighted articles should be more than 9 divisions Weighted articles should be removed and the scale goes back to zero before print out the next record 10 Press key MC CK to select external devices Options are PC AX TP Godex SH ET TDP and ZEBRA 11 Press Key MR to switch On or OFF RTC function 12 When setting is completed press key ZERO to save and retu...

Страница 11: ...ll load is 25kg 4 Wait till appears then put weights of 2 3 of full load 50kg on and press down key TARE 5 Wait till appears then put weights of full load 75kg on and press key TARE 6 The calibration Procedure is completed with a symbol of flashing then take away the weights 7 Press key TARE to return to weighing mode 6 2 Single point calibration 1 Turn on the power while pressing key SMPL to init...

Страница 12: ...ust be held in a container taring stores the container weight into memory 1 Under the weighing mode place the container on the scale weight window shown is 0 500kg Press the Key TARE to complete tare action The symbol appears and with symbol pointing at NET 2 To clear tare with an empty pan Press down the key TARE or key ZERO Preset Tare 1 Under the unload condition Enter tare value using the nume...

Страница 13: ...upper lower limits setting is blinking 2 Use numeric keys to set the upper lower Limits Key is to save upper limit value and advance to the lower limit setting 3 When setting is completed again press key MC CK to return to weighing and checking mode 4 To cancel checking action long press key MC CK then press key twice Checking action is cancelled with displays momentarily before the screen go back...

Страница 14: ...eight without the need to count additional parts A higher reference weight is important when there is a risk of inconsistent piece weights or if the reference weight is close to the minimum ACAI uses an initial averaging unit weight to count additional pieces that are placed on the scale after a few seconds the scale gives a beep as the new higher reference weight is used to recalculate the averag...

Страница 15: ... accumulation events in detail Accumulation clear To clear the total accumulation data or the first 10 accumulation events Press key MC CK while the data is displayed Symbol disappears when the total accumulation data is deleted 8 Serial Interface If external interface is needed please select the proper two in one board first which integrates RTC time display RS 232 and relay weight checking funct...

Страница 16: ...uter 8 4 RS 232 Output Format Baud Rate 2400 4800 9600 Data Bit 8 Parity N None Stop Bit 1 Code ASCI Bit Format LSB MSB 0 1 2 3 4 5 6 7 8 Start Parity Stop Data Format Kg G N W k g CR LF weight U W g p c s CR LF Unit weight T o t a l p c s CR LF pcs Example G W 2 2352 kg U W 0 5352 g pcs Total 4176 pcs lb G N W l b CR LF weight U W l b p c s CR LF Unit weight T o t a l p c s CR LF pcs Example G W ...

Страница 17: ...rvice EEPROM Check sum failure Re sold EEPROM or contact our Service The weighed articles are overload Do not load the item exceeds the maximum tolerance The calibration value exceeds the full load capacity while processing single point Calibration Adjust the calibration value The accumulated number of weighments total count or weight exceeds display range No more accumulations total count window ...

Страница 18: ...JWI 700C 17 ...

Страница 19: ...JWI 700C 18 ...

Страница 20: ...JWI 700C 19 ...

Отзывы: