INSTRUCTION MANUAL CODE 80324 REV 0
Nov/2015 (It can be modifi ed without notice) Page: 14
J
.P. SELECTA s.a.u. - Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362 e-mail: [email protected] - website: http://www.jpselecta.es
Fig 1. Programming screen
7.2 Programming screen
7.2.1 Set time and date
Press on Date/time button.
Press over push buttons to set time and date.
Use Up and Down icons to entry a value.
7.2.2 Set temperature, language and screen contrast
In a similar way set the temperature, language and screen contrast.
7.2.3 Working parameters protected under password
Some parameters settings are password protected.
Temperature calibration.
....
LANGUAGE
CONTRAST
INSERT
PASSWORD
CHANGE
PASSWORD
DATE/TIME
UNITS
CAL
7.3
Temperature and working time setting
Up to 8 programms of temperature and time can be stored.
Each program has the following parameters:
Temperature:
Temperature: from 5ºC ... 200ºC
Time:
HH:MM
Undertemp:
Alarm value for temperature drop.
Overtemp:
Alarm value for overtemperature.
From main menu press over «PROGRAMMING».
Fig7.3 screen is displayed.
Select a program (from 00 to 07) with keys:
Set working temperature and the other parameters:
Press:
keys
will be shown to select temperature.
In a similar way press:
Sets de working time.
A time setting value of 00.00 means «conti-
nuous operation».
Sets the undertemperature alarm value.
Sets the overtemperature alarm value.
Press to save .
TEMP: 16.0ºC
UNDERTMP: 0ºC
TIME: 00:00
OVERTMP: 0ºC
SAVE
PROGRAM 00
SAVE
TIME: 00:00
TEMPERATURE: 16.0ºC
OVERTEMP: 0ºC
UNDERTEMP: 0ºC
SETUP
HH:MM:SS DD/MM/YY
Fig 7.3. Programming screen.
Selected program.
Program selector.
SEL