10 Technical service
For repair and substitution of deteriorated parts, please contact with J.P. SELECTA,
s.a.u. or your distributor. They will inform you about the technical services author-
ized in your area.
Find a spare parts list for easy buying and identifying.
10.1 Spare parts
16213
Electrovalve N. A.
16067
Water electrovalve.
16025
Distilled outlet clamp valve.
24278
Water bomb support.
28129
Lockgum.
28061
Drop tray.
32050
Steam generator 2400W.
47043
Cooler glass DE-1626
47044
Rectifier glass DE-1626
47045
Cooled union tube glass DE-1626
47259
Anti feedback valve.
29520
Relay electronic PCB.
29511
Control panel electronic PCB
Hoses:
46030
TEFLON TUBE 8X10MM
Tube inmersed on the sample.
46041 Silicone tube 8x14
Water and Steam.
46042 Silicone tube 6x10
Water and Steam.
46143
3/4 WATER INLET HOSE
Connection to the water network
28264 LFL Tygon tube Distilled outlet.
32050
WARNING
Before any work inside the equipment
disconnect it from the mains.
28129
Adjusting screew
10.1.1 Adjusting the level sensor
It could be possible that along the time the collector level sensor changes it sensibility
and never detects the distilled level or it actuates just by sensing the glass flask.
Turn the adjusting screew (see picture) to adjust it.
Place a flask with water on the collector and turn the screew until the yellow LED ligths.