7.4 Selecting a program for operation
To select a stored program press over SEL
(SEL indication changes its color to yellow).
Once selected, this program in placed on the WORKING screen.
7.5 Operation screen
On the operation screen «TASK», the program selected on the programming
screen is shown (See 7.4) and it starts when pressing over .
This screen shows:
Actual temperature
Temperature set
Working time set (SET)
Elapsed time
Fig 7.4. Selecting a program
PROGRAM 04
SAVE
TIME: 00:00
TEMP: 50.0ºC
OVERTEMP: 0ºC
UNDERTEMP: 0ºC
SETUP
HH:MM:SS DD/MM/YY
SEL
SEL change to yellow
colour when programm is
selected
Fig 7.5. Operation screen
HH:MM:SS DD/MM/YY
28.9
ºC
SET
50.0
FLOW
80%
PROG 04
SET -- h -- m
T
00 h 00m
7.6 Circulation pump
The liquid circulation pump could operate in two confi guration modes:
- INTERNAL circulation:
Water (or other) will be pumped from horizontal pump output.
(See fi g 7.6.1).
Use this confi guration to made water circulation into the bath vessel in order
to homogenize the temperature from sample load.
- EXTERNAL circulation:
Water (or other) will be pumped from vertical pump output.
(See fi g 7.6.2).
Use this confi guration to made water circulation out of the bath vessel in order
to temperate an external equipment.
Fig 7.6.1 INTERNAL circulation.