GA-200 Gas Analyzer

Overview

The GA-200 Gas Analyzer uses sensors to measure and display the 
concentrations of oxygen and carbon dioxide in a sample as the percentage of a 
gas in the sample by volume. This method of expressing the concentration of a gas 
in a sample is also known as the percent volume fraction. In addition to the 
sampling and sensing systems in the GA-200, the unit has a fluorescent display 
used for observing measurements and analog outputs that allow the unit to be 
connected to iWorx data recorders.
When measuring the concentrations of oxygen and carbon dioxide in a gas, a 
sample is pumped from the input port on the front panel of the GA-200, through the 
sample cell, and out an exhaust port on the back of the unit. Gas samples are 
drawn into the unit through an external filter on the input port that protects the 
sensors from contamination and through Nafion tubing connected to the filter that 
removes moisture from the samples that would affect the measurements and the 
unit's calibration.
The GA-200 Gas Analyzer has the ability to measure and record both oxygen and 
carbon dioxide concentrations over the ranges normally recorded from human and 
animal subjects. 

Front Panel

GA-200 Front Panel

GA-200 Gas Analyzer 

Technical

Note

iWorx Systems, Inc.

www.iworx.com

LabScribe is a trademark of 

iWorx Systems, Inc. 

©2015 iWorx Systems, Inc.

Содержание GA-200 Series

Страница 1: ...oxide in a gas a sample is pumped from the input port on the front panel of the GA 200 through the sample cell and out an exhaust port on the back of the unit Gas samples are drawn into the unit throu...

Страница 2: ...to a computer through a serial RS 232 data stream The default data format is 9600 baud 8 bit and no parity with alternate baud rates of 2400 4800 19200 and 38400 The signals present are RXD TXD and g...

Страница 3: ...the 20 mA current output is set to 100 Socket 13 is the Ground for the both current outputs Socket 14 is the oxygen current output Socket 15 is the carbon dioxide current output GA 200B Rear Panel GA...

Страница 4: ...the surface of a detector The output of the sensor is inversely proportional to the concentration of oxygen in the sample because the amount of light reaching the detector decreases as the concentrati...

Страница 5: ...en the output of the carbon dioxide sensor which is located on the terminal block on the rear panel of the analyzer and the BNC input of Channel 1 on the IWX 214 Red wire into socket 15 and black wire...

Страница 6: ...lve on the regulator of the cylinder providing the second gas sample Adjust the flow rate to low Make sure the flow rate from the gas cylinder is greater than the flow rate on the gas analyzer 2 While...

Страница 7: ...cribe toolbar so that two cursors appear on the Main window Place one cursor on the section of data recorded when gas analyzer was collecting a sample of room air and the second cursor on the section...

Страница 8: ...the GA 200 gas analyzer 5 Notice that the voltages from the positions between the cursors are automatically entered into the value equations Enter the two concentrations of carbon dioxide measured fr...

Страница 9: ...be disinfected An external barrier filter on the inlet port must be used to prevent contamination of the sample circuit 4 If the GA 200 is used to collect samples of inspired and expired air during b...

Страница 10: ...zer 12VDC 1 5A Requirements of the External Power Supply 95 250VAC 47 63 Hz Dimensions 190 x 76 x 280mm WxHxD Weight Analyzer 2 3kg Power Supply 0 7kg Storing Before storing the GA 200 connect an exte...

Отзывы: