MONITOR DESCRIPTION
MONITOR DESCRIPTION
The Model 101R/NR Portable Patient Monitor is a dual trace monitor for ECG and heart rate, which produces an
output pulse corresponding in time to the R-wave. The monitor is intended primarily for use in timed imaging
studies, TMR/PMR and any application requiring precision R-wave synchronization.
Specifications
ECG
Lead
Selection:
LI,
LII, LIII menu selectable.
Patient Cable:
6-Pin AAMI Standard connector
Isolation:
Isolated from ground related circuits by >4 kV rms, 5.5 kV peak
CMRR:
≥
90 dB with patient cable and 51 k
Ω
/47 nF imbalance
Input
Impedance:
≥
20 M
Ω
at 10 Hz with patient cable
Frequency
Response
CRT:
Filtered:
0.5
to
25
Hz
Unfiltered:
0.5
to
45
Hz
Frequency
Response
X1000 output:
Filtered:
0.5 to 25 Hz
Unfiltered:
0.2
to
100
Hz
Input Bias Current:
Any lead configuration <200 nA dc maximum
Electrode
Offset
Potential:
±
0.5 V DC
Noise:
<20
µ
V peak-to-peak, referred to the input with all leads connected
through
51
k
Ω
/47 nF to ground
Defibrillator
Protection:
Protected against 360 J discharge and electrosurgery potentials
Leakage
Current:
<10
µ
A at normal condition
Electrosurgical
Interference
Protection:
Standard
Notch
Filter:
50/60
Hz
(automatic).
Model 101R/NR Service Manual
4
Содержание 101NR
Страница 1: ...Service Manual Model 101R NR Patient Monitor R wave Trigger 2409 00 20 ...
Страница 15: ...MONITOR DESCRIPTION MENU STRUCTURE HIGH LOW TRIG DELAY 101R Model 101R NR Service Manual 9 ...
Страница 56: ...BOARD LAYOUT DIAGRAMS Model 101R NR Service Manual This page is intentionally left blank ...
Страница 57: ...BOARD LAYOUT DIAGRAMS Model 101R NR Service Manual A BOARD LAYOUT DIAGRAMS ...
Страница 58: ......
Страница 59: ......
Страница 60: ......
Страница 61: ...SCHEMATICS Model 101R NR Service Manual B SCHEMATICS ...
Страница 62: ...SCHEMATICS Model 101R NR Service Manual This page is intentionally left blank ...
Страница 63: ......
Страница 64: ......