11. Specifications
General
BTS Type
4-channel
2-channel
Dimensions
440 x 350 x 178 mm
Weight
<TBD> KG
<TBD> KG
Mains voltage
200
–
250Vac (nominal single phase)
Mains frequency
50/60 Hz
Battery channels
4
2
3.5mm jack battery temperature sensor inputs
4
2
Number of internal chargers
3
2
Internal charger nominal voltages
12V/24V/combination
Output voltage range (charging)
0-18V(12V) or 0-36V(24V)
Output current range (charging)
0-20A(12V) or 0-12A(24V)
Number of internal discharge units
1
Internal discharge unit nominal voltages
12V/24V
Internal discharge unit current
0-40A
Internal discharge unit maximum power
600W
Above 35 °C ambient temperature the maximum
discharge power is derated
Battery discharge rate (C5)
20-200 Ah(User configurable)
Supported battery voltages
12V/24V
Discharge current measurement accuracy
< 1% Full Scale
Discharge voltage measurement accuracy
< 1% Full Scale
3.5mm jack serial input
1
Female USB2.0 type A port for flash drive
1
USB flash drive format
FAT32
Touchscreen
7 Inch
Firmware upgrade
Yes, using USB cable
Cooling
Active, 4 fans
Active, 3 fans
Use
For indoor use only
Ambient temperature range, active
0
–
40°C
Ambient temperature range, storage
-15 - +50°C
Maximum air humidity
95% (non condensing)
IP Class
IP20
Standardization
CE (EMC, RoHS)
Standards
EN-61326-1, EN61010-1, EN 60335-2-29