Ivoclar Vivadent Programat P300 Скачать руководство пользователя страница 28

28

Index Category Error ERR

Conti-

Error 

Message 

Text

No.

nuation
possible 

24

System

Checksum error in the 

45

Invalid checksum of the memory for firing program data - the firing program data are written to the internal

firing program memory

memory using the original values.

25

System

Write error in the 

46

Error while writing firing group data to the internal memory.

firing group memory

26

System

Read error in the 

47

Error while reading the firing group data from the internal memory.

firing group memory

27

System

Checksum error in the 

48

Invalid checksum of the memory for firing group data - the firing group data are written to the internal 

firing group memory

memory using the original values.

28

System

OT1 <> OT2 

54 

nein 

Error in the temperature measuring circuit! Contact your local Ivoclar Vivadent Service Center.

plausibility check:

**,***

OT1 = OT2 +/- 10°C

29

System

Temperature in the furnace 

56

Temperature in the furnace base is lower than 1 °C. Bring the furnace base to a higher operating temperature. 

base is lower than 1 °C

30

System 

Program start blocked

103

Starting a program is not possible due to a technical malfunction.

31

System

Incorrect time settings 

107

The clock settings are incorrect. Please set a correct date and a correct time!

(date / time)

33

Entry

HV > H (H2)

110

Enter a lower value for HV or a higher value for H (H2)

35

Entry

"Share of the holding time  120

Activate the vacuum during the holding time Tx or deactivate HV.. 

with vacuum" is activated,  
but Vx2 does not 
correspond to Tx or Tx+1

36

System

Supply voltage outside the 

700 

The supply voltage is outside the acceptable range. Check the supply voltage. 

acceptable range  

37

System

Start-up aborted due 

701

The self-test of the furnace was interrupted by an error. It is not possible to work with the furnace! Switch 

to an error 

***

the furnace off and on again, once the error has been rectified. 

38

System

Brief power failure during a 702

A firing program in progress was interrupted by a brief power failure. The program is continued!  

firing program in progress 

39

System

Power failure during a

703

A firing program in progress (started from the USB memory stick) was interrupted by a power failure.

firing program from the 

The program could not be continued, since the USB memory stick is no longer present! 

memory stick in progress – 
memory stick no longer 
present.

40

System

Prolonged power failure

704

An overnight program (firing program) in progress was interrupted by a prolonged power failure. 

during an onvernight 

The overnight program is continued!

program in progress

41

System

Reading and processing 

705

no

Error during measuring the supply voltage. Contact your local Ivoclar Vivadent Service Center.

supply voltage

**,***

42

System

Reading the power frequency 706

Error during measuring the supply voltage. Contact your local Ivoclar Vivadent Service Center. 

43

System

Incorrect supply voltage

707

The furnace is operated with the incorrect supply voltage. Make sure that the furnace is operated with the 
supply voltage indicated on the rating plate. 

44

System

Final vacuum value not 

800

The required final vacuum value cannot be reached. Check the vacuum pump. 

reached

45

System

Vacuum drop

801

An unacceptable vacuum drop has occurred.

46

System

The vacuum does not 

802

No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no 

increase (self-test)

contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?
Is the fuse F1 o.k.?

47

System

Temp. EXTERNAL T-SENSOR 1010

Temperature channel EXTERNAL T-sensor excess temperature

excess temperature 
(> 1225°C)

48

System

Write error in the furnace   1011

Error while saving furnace configuration data to the internal memory.

configuration data memory

49

System

Read error in the furnace 

1012

Error while reading the furnace configuration data from the internal memory.

configuration data memory

50

System

Checksum error in the

1013 

Invalid checksum of the memory - furnace configuration data are written into the internal memory using

furnace configuration   

the original values.

data memory

51

System

Write error in the furnace 

1014

Error while saving the furnace operational data to the internal memory.

operational data memory 

52

System 

Read error in the furnace  

1015

Error while reading the furnace operational data from the internal memory.

operational data memory  

53

System

Checksum error in the 

1016

Invalid checksum of the memory – furnace operational data are written into the internal memory using

furnace operational data 

the original values.

memory

Содержание Programat P300

Страница 1: ...Operating Instructions Programat P300...

Страница 2: ...2...

Страница 3: ...Explanation of the key functions 5 3 Program structure 5 4 Adjustable parameters and possible value ranges 5 5 Settings test programs and information 5 6 Explanation of the symbols on the display 5 7...

Страница 4: ...hose connection 23 Rubber feet 24 Protective cover vacuum 25 Furnace head housing 26 Thermocouple plug 27 Plug fuse 28 Heater plug 29 Heater plug socket 30 Thermocouple plug socket 32 Leaf spring 33 A...

Страница 5: ...5 40 25 34 6 10 35 23 47 33 1 7 5 2 3 4 8 9...

Страница 6: ...6 37 39 36 11 38 25 19 13 12 49 26 30 29 50 51 27 28 11 32 43 12 13 15 19 46 21 17 18 16...

Страница 7: ...y temperature 82 Closing time 83 Temperature increase 84 Holding temperature 85 Holding time 86 Vacuum on 87 Vacuum off 88 Long term cooling 90 Open furnace head 91 Close furnace head 92 Numeric keys...

Страница 8: ...joy working with the P300 1 2 Introduction The signs and symbols in these Operating Instructions facilitate the finding of important points and have the following meanings Risks and dangers Important...

Страница 9: ...must be observed The furnace must be operated under the indicated environmental and operating conditions Chapter 9 The Programat P300 must be properly maintained 2 1 1 Risks and dangers The furnace he...

Страница 10: ...e skin grease contamination as the parts will be prematurely dama ged 2 1 12 Risk of crushing and burn hazard Never reach under the furnace head with the hand or other parts of the body during operati...

Страница 11: ...nditions over an extended period of time Use only original spare parts The temperature range for faultless operation is 5 C to 40 C 41 F to 104 F If the furnace has been stored at very low temperature...

Страница 12: ...uum may be created with a vacuum pump The firing process is controlled with the corresponding electronic con trols and a software Moreover the set and actual temperatures are continuously compared 3 4...

Страница 13: ...ly use the two corresponding inserts Fold the side flaps 2 and combine the two packaging parts by means of the transportation flaps The packaging may be disposed with the regular household refuse 4 2...

Страница 14: ...e 5 on the firing plate holder 48 Secure the cooling tray 34 with the two screws 35 including the silicone rest 47 Place the cooling tray 34 on the frame plate 7 Make sure that the cooling tray 34 is...

Страница 15: ...ace head mounting 43 Ensure that the furnace head mounting mark 41 is aligned with the furnace base mounting mark 42 Make sure that the firing plate 5 is not damaged by mounting the furnace head Step...

Страница 16: ...itional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord 16 with the power socket 17 of the furn...

Страница 17: ...d 16 with the wall socket 2 Put the On Off switch 11 at the rear of the furnace on position I The furnace will now automatically conduct a self test The performance of all furnace components is automa...

Страница 18: ...n menu The right key can be used to navigate between the settings and auxiliary programs Cursor keys 79 80 The current cursor position is indicated by means of a fix non blinking frame around the valu...

Страница 19: ...uration and information 5 4 Adjustable parameters and possible value ranges Symbol Parameter Value range Value range P Program number P 1 100 B Stand by temperature 100 700 C 212 1292 F S Closing time...

Страница 20: ...ntered by using the numeric keys The date can be selected by using the numeric keys Settings Write protection Ivoclar Vivadent optimized temperature control function Pre drying Vacuum test program Hea...

Страница 21: ...reset to 0 Only after entry of the user code Settings Serial number Software version Furnace head firing hours Operating hours furnace Operating hours vacuum pump Latest start of calibration program...

Страница 22: ...ox 10 seconds with the beeper signal set by the user If the STOP key is not pressed during this time the beeper sounds again after 5 minutes for a period of 5 minutes Subsequently there will be no fur...

Страница 23: ...priate symbol first or second stage Two stage program stage 1 is shown stage 2 is shown If the cursor is positioned on the one stage symbol and the symbol is switched to the two stage symbol by pressi...

Страница 24: ...ttom line until the next value is entered and successfully confirmed or the process is aborted with ESC The old valid value reappears Please refer to the parameter details for the value range 6 4 2 Pr...

Страница 25: ...on the surface of the insulation are harmless and do not influence the function of the furnace in a negative fashion weekly Thermocouple 4 Check if the thermocouple is straight and in the right place...

Страница 26: ...s the ENTER key The following is displayed c Insert silver wire into the Ivoclar Vivadent sample holder d Place the sample holder with the silver strip in the centre of the firing plate 6 e Press STAR...

Страница 27: ...local Ivoclar Vivadent Service Center 13 System Heating muffle very old 23 The heating muffle is very old It is recommended to replace it After the error message has been acknowledged a firing program...

Страница 28: ...ng program from the The program could not be continued since the USB memory stick is no longer present memory stick in progress memory stick no longer present 40 System Prolonged power failure 704 An...

Страница 29: ...ure 69 System Reading calculating the 1204 no Error while measuring the furnace control temperature furnace control temperature 70 System Reading caluclating the 1205 Error while measuring the resista...

Страница 30: ...and on again If it still does not work properly contact your local Ivoclar Vivadent Service Center Vacuum pump does not start working Is the vacuum pump fuse defective Check fuse and replace if necess...

Страница 31: ...e Depth 368 mm width 303 mm 398 mm with Cooling tray Height 320 mm Usable size of the firing chamber Diameter 80 mm Height 48 mm Max firing temperature 1200 C 2192 F Weight Furnace base 12 3 kg Furnac...

Страница 32: ...local Service Center Important information The current program table is also available at www ivoclarvivadent com The program tables can be downloaded from the Internet as PDF files Please make sure t...

Страница 33: ...66 77 80 www ivoclarvivadent it Ivoclar Vivadent S A de C V Av Mazatl n No 61 Piso 2 Col Condesa 06170 M xico D F Mexico Tel 52 55 5062 1000 Fax 52 55 5062 1029 www ivoclarvivadent com mx Ivoclar Viv...

Отзывы: