Index Category Error ERR
Conti-
Error
Message
Text
No.
nuation
possible
24
System
Checksum error in the
45
Invalid checksum of the memory for firing program data - the firing program data are written to the internal
firing program memory
memory using the original values.
25
System
Write error in the
46
Error while writing firing group data to the internal memory.
firing group memory
26
System
Read error in the
47
Error while reading the firing group data from the internal memory.
firing group memory
27
System
Checksum error in the
48
Invalid checksum of the memory for firing group data - the firing group data are written to the internal
firing group memory
memory using the original values.
28
System
OT1 <> OT2
54
nein
Error in the temperature measuring circuit! Contact your local Ivoclar Vivadent Service Center.
plausibility check:
**,***
OT1 = OT2 +/- 10°C
29
System
Temperature in the furnace
56
Temperature in the furnace base is lower than 1 °C. Bring the furnace base to a higher operating temperature.
base is lower than 1 °C
30
System
Program start blocked
103
Starting a program is not possible due to a technical malfunction.
31
System
Incorrect time settings
107
The clock settings are incorrect. Please set a correct date and a correct time!
(date / time)
33
Entry
HV > H (H2)
110
Enter a lower value for HV or a higher value for H (H2)
35
Entry
"Share of the holding time 120
Activate the vacuum during the holding time Tx or deactivate HV..
with vacuum" is activated,
but Vx2 does not
correspond to Tx or Tx+1
36
System
Supply voltage outside the
700
The supply voltage is outside the acceptable range. Check the supply voltage.
acceptable range
37
System
Start-up aborted due
701
The self-test of the furnace was interrupted by an error. It is not possible to work with the furnace! Switch
to an error
***
the furnace off and on again, once the error has been rectified.
38
System
Brief power failure during a 702
A firing program in progress was interrupted by a brief power failure. The program is continued!
firing program in progress
39
System
Power failure during a
703
A firing program in progress (started from the USB memory stick) was interrupted by a power failure.
firing program from the
The program could not be continued, since the USB memory stick is no longer present!
memory stick in progress –
memory stick no longer
present.
40
System
Prolonged power failure
704
An overnight program (firing program) in progress was interrupted by a prolonged power failure.
during an onvernight
The overnight program is continued!
program in progress
41
System
Reading and processing
705
no
Error during measuring the supply voltage. Contact your local Ivoclar Vivadent Service Center.
supply voltage
**,***
42
System
Reading the power frequency 706
Error during measuring the supply voltage. Contact your local Ivoclar Vivadent Service Center.
43
System
Incorrect supply voltage
707
The furnace is operated with the incorrect supply voltage. Make sure that the furnace is operated with the
supply voltage indicated on the rating plate.
44
System
Final vacuum value not
800
The required final vacuum value cannot be reached. Check the vacuum pump.
reached
45
System
Vacuum drop
801
An unacceptable vacuum drop has occurred.
46
System
The vacuum does not
802
No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no
increase (self-test)
contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?
Is the fuse F1 o.k.?
47
System
Temp. EXTERNAL T-SENSOR 1010
Temperature channel EXTERNAL T-sensor excess temperature
excess temperature
(> 1225°C)
48
System
Write error in the furnace 1011
Error while saving furnace configuration data to the internal memory.
configuration data memory
49
System
Read error in the furnace
1012
Error while reading the furnace configuration data from the internal memory.
configuration data memory
50
System
Checksum error in the
1013
Invalid checksum of the memory - furnace configuration data are written into the internal memory using
furnace configuration
the original values.
data memory
51
System
Write error in the furnace
1014
Error while saving the furnace operational data to the internal memory.
operational data memory
52
System
Read error in the furnace
1015
Error while reading the furnace operational data from the internal memory.
operational data memory
53
System
Checksum error in the
1016
Invalid checksum of the memory – furnace operational data are written into the internal memory using
furnace operational data
the original values.
memory