24

6.3 Firing using an individual program

Step 1:

Switch to the program selection by pressing the Home key (93).
Select Individual programs with the right cursor key. Select a free
program.

Step 2:

Press the desired parameter (e.g.
temperature increase). The input
mask appears. In the lower field,
you will find information about the
acceptable temperature range (min.
and max.). This will facilitate the
entry and prevents input errors.

Step 3:

Press the START key (73) to start
the program. The process is
indicated in the firing curve display.

6.4 Further possibilities and special features of the

furnace

6.4.1 Illogical values or incorrect entry

If an illogical value is entered by means of the numeric keys
(outside the current value range), the invalid entry still blinks after
confirmation.

As error message (entry error: entry
outside the value range), an excla-
mation mark blinks in the bottom
line until the next value is entered
and successfully confirmed or the
process is aborted with ESC. The
old, valid value reappears. Please
refer to the parameter details for the value range.

6.4.2 Program write protection

If the program write-protection is activated in the Settings, all Ivoclar
Vivadent programs are write-protected. This is indicated by a closed-
lock icon in the stand-by display and the parameter list.

Only Ivoclar Vivadent programs can be protected with the

program write-protection. This function is not available for
individual programs.

6.4.3 Stopping the running program

Press the STOP key once to pause a running program. The green LED
in the START key blinks. Press the STOP key twice to completely stop
the program or press START to continue. The current temperature is
maintained.

6.4.4 Changing the parameters while the program is running

All parameters of the program which have not yet been executed,
can be changed while the program is stopped (green LED blinks).

6.4.5 Quick opening of the furnace head

By switching the symbol “standard opening of the furnace head”
for the currently loaded program, the user may activate a ”quick
opening of the furnace head” at the end of the program.

6.4.6 Closing of the furnace head

Various functions are available for the
closing of the furnace head. If the cursor
is on the spot in the parameter list shown
in the figure, you can toggle between the
following functions by using the + or –
key:

Thermo Shock Protection

Pre-drying

no function activated

– TSP – Thermo Shock Protection

The TSP function prevents the object from too high of
temperatures during the closing process. For this purpose, the TSP
function gauges the temperature of the firing chamber in the fur-
nace head upon start of the firing program. If required, the closing
path within the set closing time S is adjusted. TSP is only active if
no active predrying has been selected or if the temperature in the
firing chamber is too high when the program is started. In
addition, TSP works only if the stand-by temperature of
B = 403 °C/757 °F required for Ivoclar Vivadent materials is used.
The active TSP function is shown on the display by means of the
symbol "TSP". If programs are started with a temperature of more
than 680 °C / 1256 °F, an error message is produced.

– Pre-drying

If the pre-drying function is activated, the furnace heats to the
specified stand-by temperature.

6.4.7 Power-save mode

If the power-save mode is activated and the furnace head closed,

this function is automatically activated after 30 minutes if the fur-
nace is idle and no key is pressed during this time. The Power Saving
icon appears on the display. The power-save mode is terminated by
pressing any key.

6.4.8 Protocolling

If the protocol function is activated (see section 5.5.1), a firing

protocol can be copied to a PC at the end of the firing program by
means of the included software. Subsequently, the protocols can be
edited and managed on the PC.

The furnace saves the 15 firing programs that were last
conducted for protocolling. Copying the protocols to a PC
will automatically erase the saved entries on the furnace.

6.4.9 Software update

The user will be able to conduct a software update by PC and down-
load cable. For that purpose, the software download mode of the
furnace is activated by pressing two special keys simultaneously
while the power supply is switched on. For further details, please
refer to the Software Update Instructions
(www.ivoclarvivadent.com/downloadcenter).

Please note that the individual programs are being erased
upon a software update. We therefore recommend that
you carry out a data backup (e.g. using the Programat
PrograBase 2 PC software).

Содержание Programat P300/G2

Страница 1: ...Operating Instructions Programat P300 G2 s Valid as of software Version 5 0...

Страница 2: ...2...

Страница 3: ...ey functions 5 3 Program structure 5 4 Adjustable parameters and possible value ranges 5 5 Settings test programs and information 5 6 Explanation of the symbols on the display 5 7 Explanation of the b...

Страница 4: ...use 28 Heater plug 29 Heater plug socket 30 Thermocouple plug socket 32 Leaf spring 33 Air vents base 34 Cooling tray 35 Screw for cooling tray 36 Hood 37 Knurled screw for hood 38 Air vents furnace h...

Страница 5: ...5 40 25 34 6 10 35 23 47 33 1 7 5 2 3 4 8 9...

Страница 6: ...6 49 26 30 29 50 51 27 28 11 32 43 37 39 36 11 38 25 19 53 13 12 12 13 15 19 46 21 17 18 16...

Страница 7: ...losing time 83 Temperature increase 84 Holding temperature 85 Holding time 86 Vacuum on 87 Vacuum off 88 Long term cooling 89 Power saving key 90 Open furnace head 91 Close furnace head 92 Numeric key...

Страница 8: ...cation Burn hazard Risk of crushing The Operating Instructions must be read 1 3 Notes regarding the Operating Instructions Furnace concerned Programat P300 G2 Target group Dental technologists These O...

Страница 9: ...e must be observed The furnace must be operated under the indicated environmental and operating conditions Chapter 9 The Programat P300 G2 must be properly maintained 2 1 1 Risks and dangers The furna...

Страница 10: ...l be prematurely damaged 2 1 11 Risk of crushing and burn hazard Never reach under the furnace head with the hand or other parts of the body during operation since there is a risk of crushing and a bu...

Страница 11: ...original spare parts The temperature range for faultless operation is 5 C to 40 C 41 F to 104 F If the furnace has been stored at very low temperatures or high atmospheric humidity the head has to be...

Страница 12: ...hat a vacuum may be created with a vacuum pump The firing process is controlled with the corresponding electronic con trols and a software Moreover the set and actual temperatures are continuously com...

Страница 13: ...mply use the corresponding inserts Fold the side flaps 2 and combine the two packaging parts by means of the transportation flaps The packaging may be disposed of with the regular household refuse 4 2...

Страница 14: ...5 on the firing plate holder 48 Secure the cooling tray 34 with the two screws 35 including the silicone washer 47 Place the cooling tray 34 on the frame plate 7 Make sure that the cooling tray 34 is...

Страница 15: ...ace head mounting 43 Ensure that the furnace head mounting mark 41 is aligned with the furnace base mounting mark 42 Make sure that the firing plate 5 is not damaged by mounting the furnace head Step...

Страница 16: ...ng additional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord 16 with the power socket 17 of th...

Страница 17: ...ion I The furnace will now automatically conduct a self test The performance of all furnace components is automatically checked The display shows the following indications during the self test 1 Statu...

Страница 18: ...ix non blinking frame around the value Enter keys 76 77 92 The values can be entered by using the keys or the numeric keys Each value entered by means of the or key will be immediately accepted as lon...

Страница 19: ...ne or two stage program by using the or key 5 4 Adjustable parameters and possible value ranges Symbol Parameter Value range Value range P Program number P 1 300 B Stand by temperature 100 700 C 212 1...

Страница 20: ...ntered by using the numeric keys The date can be selected by using the numeric keys Settings Write protection Ivoclar Vivadent optimized temperature control function Power saving mode Service interval...

Страница 21: ...user code The firing hour count can be reset to 0 Only after entry of the user code The operating hour count of the vacuum pump can be reset to 0 Only after entry of the user code Settings Serial num...

Страница 22: ...lf test is completed To inform the user that the automatic self test has been successfully completed the selected melody is played 2 Furnace head open and temperature below 550 C 1022 F To inform the...

Страница 23: ...being displayed Special variant Conducting a one stage program in which only part of the holding time is fired with a vacuum requirement holding temperature 1 holding temperature 2 6 1 3 Operating ind...

Страница 24: ...of the furnace head If the cursor is on the spot in the parameter list shown in the figure you can toggle between the following functions by using the or key Thermo Shock Protection Pre drying no func...

Страница 25: ...s and do not influence the function of the furnace in a negative fashion weekly monthly weekly When Thermocouple 4 Check if the thermocouple is straight and in the right place Opening mechanism Check...

Страница 26: ...s menu by pressing the Information Settings key Subsequently press and hold the right cursor key 79 until the following display appears To select the calibration pro gram press the ENTER key The follo...

Страница 27: ...to start a firing program firing program 27 no Furnace head cannot The furnace head cannot be moved to the final position It might be blocked by an external mechanical be initialized source If not co...

Страница 28: ...e following error messages is being displayed 25 29 43 44 45 46 47 48 54 56 103 107 143 144 145 146 147 148 700 701 703 704 705 706 707 1010 1011 1012 1013 1014 1015 1016 1024 1025 1026 1028 1143 1144...

Страница 29: ...program still running Wait until the program is complete Switch furnace off and on again If it still does not work properly contact your local Ivoclar Vivadent Service Centre Vacuum pump does not sta...

Страница 30: ...he closed furnace Depth 430 mm width 305 mm 410 mm with Cooling tray Height 320 mm Usable size of the firing chamber Diameter 80 mm Height 48 mm Max firing temperature 1200 C 2192 F Weight Furnace bas...

Страница 31: ...rvice Centre Important information The current program table is also available at www ivoclarvivadent com downloadcenter The program tables can be downloaded from the internet as PDF files Please make...

Страница 32: ...32 10 2 Menu structure 10 2 1 Possibilities of the program selection...

Страница 33: ...33 10 2 2 Information Settings...

Страница 34: ...34 10 3 Firing curves 10 3 1 Vacuum during both stages 10 3 2 Vacuum during long term cooling 10 3 3 Pre drying 10 3 4 Controlled long term cooling...

Страница 35: ......

Страница 36: ...7 www ivoclarvivadent com Ivoclar Vivadent S A de C V Av Mazatl n No 61 Piso 2 Col Condesa 06170 M xico D F Mexico Tel 52 55 5062 1000 Fax 52 55 5062 1029 www ivoclarvivadent com mx Ivoclar Vivadent L...

Отзывы: