ITT NEO-DYN 100T Скачать руководство пользователя страница 12

 

Enclosure 6N Temperature switches

Page 9

  

Wetted Materials 

 

Table 1 

No 

Description 

3  300-series CRES with Viton seal, except Teflon seal on Range J (Model 100T) 
5  300-series CRES with graphite-filled non-asbestos packing (Model 100TC) 

Weather Proof Ratings 

Ingress protection IP66 

Enclosure type 3, 3R, 3S, 4, 4X, 13 (NEMA) 

Pressure Ratings 

Maximum Operating Pressure: 1500 psig (103.5 bar) 

Proof Pressure: 2250 psig (155.1 bar) 

Electrical and Ambient Temperature Ratings 

Minimum ambient temperature is -40 °C (-40 °F),  

Form C3X (SPDT): 

15 amp @ 125 or 250 VAC , 1/8 HP @ 125 VAC, 

¼ HP @ 250 VAC, to +70 °C (+158 °F) max; 

5 amp resistive , 3 amp inductive @ 28 VDC; .5 amp resistive , .04 amp inductive @ 125 

VDC to +85 °C (+186 °F) max. 

Form CC3X (DPDT): 

11 amp and ¼ hp @ 125 or 250 VAC to +65 °C (+149 °F) max; 

5 amp resistive , 3 amp inductive @ 28 VDC; .5 amp resistive @ 125 VDC to+85 °C 

(+186 °F) max. 

Form C3MX or CC3MX (SPDT or DPDT with M Option): 

1 amp @ 125 VAC to +85 °C (+186 °F) max. 

Form Z3X (SPDT): 

15 amp @ 125 or 250 VAC , 1/4 HP @ 125 VAC, 

1/2 HP @ 250 VAC, to +80 °C (+176 °F) max; 

5 amp resistive , 3 amp inductive @ 28 VDC; 1 amp resistive , .5 amp inductive @ 125 

VDC to +85 °C (+186 °F) max

 

 

Содержание NEO-DYN 100T

Страница 1: ...NEO DYN MODELS 100T AND 100TC ADJUSTABLE TEMPERATURE SWITCHES ENCLOSURE 3X FOR WEATHERPROOF APPLICATIONS INSTALLATION AND OPERATION MANUAL PN 610 0012 Rev C...

Страница 2: ...al injury death or substantial property damage if the warning is ignored CAUTION Indicates a hazard which will or can cause minor personal injury or property damage if the caution is ignored SPECIAL C...

Страница 3: ...FIGURATIONS 2 CHAPTER 2 INSTALLATION 3 MOUNTING 3 PROCESS CONNECTIONS 4 PROCESS MEDIA 4 POTENTIALLY EXPLOSIVE ATMOSPHERES HAZARDOUS LOCATIONS 4 ELECTRICAL CONNECTIONS 5 CHAPTER 3 ADJUSTMENTS AFTER INS...

Страница 4: ...Model 100TC has a graphite filled non asbestos packing Setpoint adjustments are easily made through the adjustable range by turning an adjustment nut located inside the housing The standard configurat...

Страница 5: ...n or our representatives in several ways The Neo Dyn customer service phone number is 661 295 4000 Our customer service department is open from 7 00 am to 3 30 p m Pacific Time Our Internet site is ww...

Страница 6: ...MOUNTING Model 100T can be mounted directly to the process connection if there is no significant vibration and the fluid lines are capable of supporting the weight It can also be attached to a flat su...

Страница 7: ...ling compound CAUTION Avoid excessive torque on all threaded connections WARNING Do not exceed the marked maximum operating pressure in normal operation WARNING Do not use as a safety device The marke...

Страница 8: ...uirements of the NEC or applicable local or national electrical codes including wire gauges and insulation temperature ratings Conduit seals may be required Figure 2 Form C and CC Schematics Shown Bel...

Страница 9: ...below 4 To check the increasing setpoint of a temperature switch insert the temperature probe into a fluid bath so that the hex is within 3 8 inch 9 5 mm of the surface and also insert a calibrated th...

Страница 10: ...ng can be heard to audibly snap but an ohmmeter indicates no electrical switching the cause is probably stuck or burned switch contacts or the switch element has moved away from the position where it...

Страница 11: ...ls CE Mark All models with option letter X in the part number Internal switching elements are recognized components by both UL and CSA Part Number The part number contains information about which conf...

Страница 12: ...rature Ratings Minimum ambient temperature is 40 C 40 F Form C3X SPDT 15 amp 125 or 250 VAC 1 8 HP 125 VAC HP 250 VAC to 70 C 158 F max 5 amp resistive 3 amp inductive 28 VDC 5 amp resistive 04 amp in...

Страница 13: ...rd switches are listed below These options can be combined every included option will be indicated in the part number of the unit Table 2 Option Description A Epoxy painted exterior M Gold Electrical...

Страница 14: ...hipped prepaid C Exclusion from Warranty THE FOREGOING WARRANTY IS IN LIEU OF AND EXCLUDES ALL OTHER EXPRESSED OR IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS OR OTHERWISE Components manufactured...

Отзывы: