Calibration Commands
Copyright © Itech Electronic Co., Ltd.
58
CALibrate:CURRent:LIMit:POINt
This command is used to calibrate the current limit point in CV mode.
Command syntax:
CALibrate:CURRent:LIMit:POINt <point>
Parameter:
P1|P2
Unit:
None
CALibrate:CURRent:LIMit
This command is used to calibrate the current limit value in CV mode.
Command syntax:
CALibrate:CURRent:LIMit[:LEVel] <NRf>
Parameter:
Current< external reading >
Unit:
A