6
Configuration
pH
7.60
SP
7.00
Out 1 021%
Out 2 070%
m3h 10.
5
CONFIG 21.3ºC
pH
7.60
SP
7.00
Out 1 021%
Out 2 070%
m3h 10.
5
CONFIG 21.3ºC
PH
m3h
10.5
7.6
0
1
0
1
ACID
ALK.
PH
m3h
10.5
7.6
0
1
0
1
ACID
ALK.
4. OPERATION
4.1 Main screen and navigation
Confirm
Exit without
confirming
Increase / decrease value,
and move up /down
Move left / right
ENT
ENT
ESC
Regulating
output 1
Move the cursor to select the following menu:
pH, m3h or CONFIG, and enter with ENTER
Warnings on the main screen:
Flashing reading: pH alarm
Flashing regulation output: no device detected on the 4-20 mA output
Flashing regulation outputs (both): remote control deactivated, or no
device connected to the 4-20 mA outputs
Press ENTER to edit the setpoint
or move the cursor to enter the
following menu:
CAL: Calibration
CONTROL: control output
ALARM: alarms
REGISTER: output for register
pH regulation
output
Flow rate reading
Access to the Configuration menu
Calibration menu
Alarm menu
Control menu
Register menu
Cursor
Regulating
output 2
pH
CAL
CONTROL
ALARM
REGISTER
SP
07.00
PH CAL.
Point 1
07.00
Point 2 04.00
Reading 07.23
PH CONTROL
CONTROL
PI CONTROL
PH ALARM
Rearm:
Yes; No
Stop: pH;
All; None
pH +
1.00
025
s
pH -
0.50
060
s
PH REGISTER
Out:
4-20 mA
2
4 mA:
00.00
20 mA:
14.00