EXCL. CYC. MIN.
The software will exclude at least this number of cycles from
data before performing analysis.
AUTO EXCL. CYC.
By default the software will automatically determine which
cycles at the beginning of a run exhibit artefacts, and exclude
this number of cycles from data before performing analysis.
This feature can be disabled by deselecting this option.
ARTEFACT FILTERING
Biochemical and physical factors can cause fluorescence levels
to change during a run. These are often observed during early
cycles. Filtering out these artefacts can improve the accuracy
of amplification analysis. Two settings enable you to control the
process of filtering out such artefacts.