SR4-Set Setpoint Ref4
Return Value : OK
Parameter : 1 ~ 3
EX)
<Host>SR4 2<CR>
<IV2300>OK
It is setting SP4 ref by CGA
SR5-Set Setpoint Ref5
Return Value : OK
Parameter : 1 ~ 3
EX)
<Host>SR5 2<CR>
<IV2300>OK
It is setting SP5 ref by CGA
SR6-Set Setpoint Ref6
Return Value : OK
Parameter : 1 ~ 3
EX)
<Host>SR6 2<CR>
<IV2300>OK
It is setting SP6 ref by CGA
SRA-Set Ref All(SR1 ~ SR6)
Return Value : OK
Parameter : 1 ~ 3 & 6Byte Data
Parameter means that Ref Setting to *1:IG, 2:CGA, 3:CGB
and Each 1Byte indicate one SP Ref Channel
*If CM Card is used instead IG Card, 1 is CM.
EX)
<Host>SRA 112233<CR>
<IV2300>OK
It is setting SP1 & 2 Ref by IG, SP3 & 4 Ref by CGA
SP5 & 6 Ref by CGB. Its like SR1 ~ SR6
33
Содержание IVC2300
Страница 4: ...Customer Support SENSER IVC2300 Vacuum Gauge Controller 608 2 302 2...
Страница 10: ...Connecting Cables Gauge Mounting PC RS232 IVC2300 VC120 VC120 CM IG Figure 1 Connecting Cable Gauge Mounting 8...
Страница 11: ...Installation Environment Operating Environmental Requirements 15 50 59 122 Chassis Grounding 9...
Страница 27: ...Analog Output Formula IG Vout V log P 12 P 10 V 12 CG Vout V log P 7 P 10 V 7 25...