CIS-CAL INIT SYSTEM
Return Value : OK
Parameter : <None>
EX)
<Host>CIS<CR>
<IV2300>OK
It is System Initialization by Factory setting.
!CAUTION - If U use this setting, may not work right.
SR1-Set Setpoint Ref1
Return Value : OK
Parameter : 1 ~ 3
Parameter means that Ref Setting to *1:IG, 2:CGA, 3:CGB
*If CM Card is used instead IG Card, 1 is CM.
EX)
<Host>SR1 2<CR>
<IV2300>OK
It is setting SP1 ref by CGA
SR2-Set Setpoint Ref2
Return Value : OK
Parameter : 1 ~ 3
EX)
<Host>SR2 1<CR>
<IV2300>OK
It is setting SP2 ref by IG
SR3-Set Setpoint Ref3
Return Value : OK
Parameter : 1 ~ 3
EX)
<Host>SR3 3<CR>
<IV2300>OK
It is setting SP3 ref by CGB